Riester ri-former Скачать руководство пользователя страница 19

19

Screw the specula clockwise until noticeable resistance is felt. To re-

move the specula, screw the specula counter clockwise.

L3 otoscope:

Fit the chosen specula on the chrome-plated metal fixture of the oto-

scope until it locks into place. To remove the specula, press the blue 

ejection button. The specula is automatically ejected.

 

6.1.3 Swivel lens for magnification

The swivel lens is fixed to the device and can be swivelled 360°.

6.1.4 Insertion of external instruments into the ear 

 

If you wish to insert external instruments into the ear (e.g. tweezers), 

you have to rotate the swivel lens (approx. 3-fold magnification) located 

on the otoscope head by 180°. Now you can use the operation lens.

6.1.5 Pneumatic test

To perform the pneumatic test (= examination of the eardrum), you re-

quire a ball, which is not included in the normal delivery package, but 

can be ordered separately. The tube for the ball is attached to the con-

nector. Now you can carefully insert the necessary volume of air into 

the ear canal.

6.1.6 Technical data of the lamp

Otoscope XL 2.5 V 2.5 V 750 mA ave. life 15 h

Otoscope XL 3.5 V 3.5 V 720 mA ave. life 15 h

Otoscope LED 2.5 V 2.5 V 280 mA ave. life 10.000 h

Otoscope LED 3.5 V 3.5 V 280 mA ave. life 10.000 h

6.2  ri-scope®L ophthalmoscopes

6.2.1 Intended use

The Riester ophthalmoscope described in these Operating Instructions 

is produced for the examination of the eye and the eyeground. 

6.2.2 Lens wheel with correction lens

The correction lens can be adjusted on the lens wheel. The following 

correction lenses are available:

L1 and L2 ophthalmoscopes:

Plus: 1-10, 12, 15, 20, 40.

Minus: 1-10, 15, 20, 25, 30, 35.

L3 ophthalmoscope:

Plus: 1-45 in single steps

Minus: 1-44 in single steps

The values can be read off in the illuminated field of view. Plus values 

are displayed in green numbers, minus values with red numbers.

Attention!

Because  prolonged  intense  exposure  to  light  can  damage 

the retina, the use of the eye exam device should not be un-

necessarily  prolonged,  and  the  brightness  setting  should 

not be set higher than needed for a clear representation of 

the target structures. 

The irradiation dose of the photochemical exposure to the 

retina is the product of irradiance and duration of irradia-

tion. If the irradiance is reduced by half, the irradiation time 

may be twice as long to reach the maximum limit.

Although no acute optical radiation hazards have been iden-

tified  for  direct  or  indirect  ophthalmoscopes,  it  is  recom-

mended that the intensity of light directed into the patient‘s 

eye be reduced to the minimum required for examination / 

diagnosis. Infants / children, aphasics and people with eye 

diseases are at a higher risk. The risk may be increased if 

the patient has already been examined with this or another 

ophthalmological instrument during the last 24 hours. This 

is especially true when the eye has been exposed to retinal 

photography.

The  light  of  this  instrument  may  be  harmful.  The  risk  of 

eye  damage  increases  with  the  duration  of  irradiation.  An 

irradiation period with this instrument at maximum inten-

sity of longer than >5 min. exceeds the guideline value for 

hazards.

This instrument does not pose a photobiological hazard ac-

cording to DIN EN 62471 but still features a safety shutdown 

after 2 / 3 minutes.

6.2.3 Apertures

The following apertures can be selected with the aperture hand-wheel:

L1 ophthalmoscope:

Semi-circle, small/medium/large circular aperture, fixation star, slit.

L2 ophthalmoscope:

Semi-circle, small/medium/large circular aperture, fixation star and slit.

L3 ophthalmoscope:

Semi-circle,  small/medium/large  circular  aperture,  fixation  star,  slit  and 

grid. 

 

 

Aperture    

Function 

  

 

  

Semicircle:   

for examinations with turbid lenses  

        

Small circle: 

to reduce reflections for small pupils  

  

   

Medium circle: 

to reduce reflections for small pupils  

 

  

Large circle:   

for normal examination results  

 

  

Grid: 

 

for topographic determination of retina 

 

 

 

 

changes  

         

Light slit:   

to determine differences in level  

 

  

Fixation star:   

to ascertain central of eccentric fixation

6.2.4 Filters

Using  the  filter  wheel,  the  following  filters  can  be  switched  for  each 

aperture:

L1 ophthalmoscope Red-free filter

L2 ophthalmoscope Red-free filter, blue filter and polarisation filter.

L3 ophthalmoscope Red-free filter, blue filter and polarisation filter.

Filter  

 

Function

Red-free filter:  

contrast enhancing to assess fine vascular changes,  

 

 

 

e.g. retinal bleeding

Polarisation filter:  for precise assessment of tissue colours and to avoid  

 

 

 

retinal reflections

Blue filter:   

for improved recognition of vascular abnormalities or 

 

 

 

bleeding, for fluorescence ophthalmology

For L2 + L3, every filter can be switched to every aperture.

6.2.5 Focussing device (only with L3)

Fast fine adjustment of the examination area to be observed is achieved 

from various distances by turning the focussing wheel.

6.2.6 Magnifying glass

A magnifying glass with 5-fold magnification is supplied with the opht-

halmoscope set. This can be positioned between the instrument head 

and the area under examination, as required. The area under examina-

tion is magnified accordingly.

6.2.7 Technical data on the lamp

XL 2.5 V ophthalmoscope: 750 mA ave. life 15 h

XL 3.5 V ophthalmoscope: 690 mA ave. life 15 h

LED 3.5 V ophthalmoscope: 280 mA ave. life 10.000 h  

6.3  Slit and spot retinoscopes

6.3.1  Intended use

The  slit/spot  retinoscopes  (also  known  as  skiascopes)  described  in 

these Operating Instructions are produced to determine the refraction 

(ametropias) of the eye.

6.3.2 Commissioning and function

Position the required instrument head on point of attachment on top 

section  of  handle  with  both  recesses  of  the  instrument  head  bottom 

section being congruent with the two projecting guide cams of the bat-

tery handle. Press instrument head lightly on battery handle and rotate 

handle in clockwise direction to the stop. Remove head by rotating in 

counter-clockwise  direction.  Rotation  and  focusing  of  the  slit  and/or 

spot image may now be effected by the knurled screw.

Содержание ri-former

Страница 1: ...1 ri former Gebrauchsanweisung Instruction for use Instructions d utilisation Instrucciones de uso Istruzioni per l uso...

Страница 2: ...ie haben eine hochwertige Riester ri former Diagnosestation erwor ben die entsprechend der Richtlinie EU 2017 745 hergestellt wurde und st ndig strengsten Qualit tskontrollen unterliegt Bitte lesen Si...

Страница 3: ...tientenpopulation Das Ger t ist f r den Einsatz an allen Patienten vorgesehen 1 4 4 Bestimmungsgem e Anwender Benutzer Das Ger t kann von rzten und Pflegepersonal in Krankenh usern medizinischen Einri...

Страница 4: ...ebrauch die Integrit t und Vollst ndigkeit der ri former Diagnosestation des Erweiterungsmoduls der Instrumentenk pfe zu berpr fen Alle Komponenten m ssen miteinander kompatibel sein Inkompatible Komp...

Страница 5: ...Minuten ein 3 3 2 Entnehmen Sie den Handgriff nach oben aus der Griffhalterung und befestigen Sie den gew nschten Instrumentenkopf indem Sie die bei den hervorstehenden F hrungsnocken auf den Handgrif...

Страница 6: ...r t Medizinisches Ger t f r die Stromversor gung von Instrumenten Elektrischer Schutz Isolationsausr stung der Klasse II Modell ri former Diagnosestation mit ri former Erweiterungsmodul Netzteil Eing...

Страница 7: ...n der Patient bereits inner halb der letzten 24 Stunden mit diesem oder einem anderen ophthalmologischen Instrument untersucht wurde Dies gilt insbesondere dann wenn das Auge einer Netzhautfotogra fie...

Страница 8: ...abstehenden F hrungsnocken auf der Oberseite des Batteriegriffs aufsitzen Dr cken Sie den Instrumen tenkopf leicht auf den Batteriegriff und drehen Sie den Handgriff im Uhrzeigersinn bis zum Anschlag...

Страница 9: ...st und setzen Sie die Trichteraufnahme wieder auf 7 1 Otoskope L2 L3 ri derma Lampentr ger Nasenspekulum und Spatelhalter Entfernen Sie den Instrumentenkopf vom Batteriegriff Die Lampe be findet sich...

Страница 10: ...h ist sollte das ME Ger t beobachtet werden um seinen bestimmungsgem en Betrieb in dieser Anordnung zu berpr fen Dieses ME Ger t ist ausschlie lich f r den Gebrauch durch medizinisches Fachpersonal vo...

Страница 11: ...ngs stufe Elektromagnetische Umgebung Leitlinien HF Leitung IEC 61000 4 6 HF Strahlung IEC 61000 4 3 Umgebungsfelder von drahtlosen Funkkommunika tionsger ten Erf llt 3 V m 80 MHz bis 2 7 GHz Erf llt...

Страница 12: ...Stk ri former Instrumentenk pfe ohne Diebstahlsicherung mit Diebstahlsicherung ri scope L F O Otoskop Artikel Nr 10563 Artikel Nr 10563 301 L1 3 5 V XL L1 3 5 V XL mit Diebstahlsicherung Artikel Nr 10...

Страница 13: ...fter Handhabung Alle defekten Teile des Produkts werden inner halb der Garantiezeit kostenlos ersetzt oder repariert Dies gilt nicht f r Ver schlei teile F r das R1 Blutdruckmessger t sto fest gew hre...

Страница 14: ...tion 4 2 Cleaning and Disinfection 5 Technical Data 6 ri scope L Instrument heads ri scope Instrument heads 6 1 ri scope L otoscope 6 2 ri scope L ophthalmoscopes 6 3 Slit and spot retinoscopes 6 4 De...

Страница 15: ...4 3 Intended patient population The device is intended for all patients 1 4 4 Intended operators users The device could be used by a doctor nurse in hospitals medical facili ties clinics doctors offi...

Страница 16: ...All components must be compatible with each other Incompatible components can result in degraded perfor mance Never knowingly use a defective device Immediately replace parts that are broken worn miss...

Страница 17: ...bled ri former anesthesia device on the universal clamp and insert it Please ensure that both pins are introduced in the universal clamp Afterwards tighten the ri former anesthesia device with the lat...

Страница 18: ...t 2 2 x 5 V dc 2 x 1 15 A ri former Extension Module Input 5 V DC 3 A 15 W Output 1 1 x 3 5 V dc 700 mA Output 2 1 x 5 V dc 1 x 1 15 A Classification Application part type B Operating conditions 0 C t...

Страница 19: ...otography The light of this instrument may be harmful The risk of eye damage increases with the duration of irradiation An irradiation period with this instrument at maximum inten sity of longer than...

Страница 20: ...se direction For two modes of operation Fast expansion Push set screw on instrument head down with your thumb This setting does not allow changes in the position of the specula legs Individual expansi...

Страница 21: ...mp cloth until visual cleanliness is achieved Wipe with disinfectant according to the instructions of the disinfectant manufacturer Only cleaning agents with proven efficacy should be used under consi...

Страница 22: ...tion ri former should assure that it is used in such an environment Emission test Compliance Electromagnetic en vironment guidance RF Emissionen CISPR 11 Group 1 The ri former Dia gnostic Station uses...

Страница 23: ...or relocating the ri former Diagnostic Station b Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m Recommended separation distances between portable and mobile RF c...

Страница 24: ...teVue head single LED 3 5 V with anti theft device for ri former wall station Art No 10513 EliteVue head single XL 3 5 V Art No 10513 301 EliteVue head single XL 3 5 V with anti theft device for ri fo...

Страница 25: ...Donn es techniques 6 T tes d instrument ri scope L T tes d instrument ri scope 6 1 Otoscope ri scope L 6 2 Ophtalmoscopes ri scope L 6 3 R tinoscopes fente et spot 6 4 Dermatoscope 6 5 Illuminateur br...

Страница 26: ...reil est destin tous les patients 1 4 4 Op rateurs utilisateurs pr vus L appareil peut tre utilis par un m decin ou une infirmi re dans des h pitaux des tablissements m dicaux des cliniques ou des cab...

Страница 27: ...rit et l int gralit de la station de diagnostic ri former du module d exten sion de la t te d instrument Tous les composants doi vent tre compatibles entre eux L incompatibilit des composants peut ent...

Страница 28: ...is de blocage Placez l appareil ri former anesth sie pr assembl sur la pince universelle et ins rez le Veuillez vous assurer que les deux broches sont introduites dans la pince universelle Serrez ensu...

Страница 29: ...60 Hz 0 6 A Sortie 5 V CC 3 A 15 W Station de diagnostic ri former Entr e 5 V CC 3 A 15 W Sortie 1 1 x 3 5 V CC 700 mA Sortie 2 2 x 5 V CC 2 x 1 15 A Module d extension ri former Entr e 5 V CC 3 A 15...

Страница 30: ...ologique au cours des 24 derni res heures Cela est particuli rement vrai lorsque l il a t expos une photographie r tinienne La lumi re de cet instrument peut tre dangereuse Le ris que de l sions ocula...

Страница 31: ...m illuminator LED 3 5 V 3 5 V 280 mA ave life 10 000 h 6 6 Sp culums nasaux 6 6 1 Utilisation pr vue Le sp culum nasal d crit dans ce mode d emploi est con u pour l c lairage et donc l examen de l int...

Страница 32: ...s d entretien 8 1 Remarque g n rale Le nettoyage et la d sinfection des instruments m dicaux servent prot ger le patient l utilisateur et les tiers et conserver les instru ments m dicaux en bon tat En...

Страница 33: ...i dessous Le client ou l utilisateur de la station de diagno stic ri former doit s assurer que celle ci est utilis e dans un tel environnement Directives et d claration du fabricant missions lectromag...

Страница 34: ...es telles que la r orientation ou le d placement de la station de diagnostic ri former b Sur la plage de fr quence de 150 kHz 80 MHz les intensit s de champ doivent tre inf rieures 3 V m Distances de...

Страница 35: ...eVue simple XL 3 5 V Art n 10513 301 T te EliteVue simple XL 3 5 V avec dispositif antivol pour la station murale ri former 12 limination Attention Les dispositifs m dicaux utilis s doivent tre jet s...

Страница 36: ...fecci n 5 Datos t cnicos 6 Cabezales de instrumento ri scope L Cabezales de instrumento ri scope 6 1 Otoscopio ri scope L 6 2 Oftalmoscopio ri scope L 6 3 Retinoscopios de hendidura y punto 6 4 Dermat...

Страница 37: ...os los pacientes 1 4 4 Operadores usuarios previstos El dispositivo puede ser utilizado por personal m dico o de enfermer a en hospitales centros m dicos cl nicas o consultorios m dicos No se puede us...

Страница 38: ...usuario Es su responsabilidad Antes de cada uso el usuario debe verificar la integridad y la totalidad de la unidad de diagn stico ri former m du lo de extensi n cabezal de instrumentos Todos los com...

Страница 39: ...tivo de an estesia ri former premontado en la abrazadera universal e ins rtelo Aseg rese de que ambos pasadores se introducen en la abrazadera universal Luego apriete el dispositivo de anestesia ri fo...

Страница 40: ...1 x 3 5 Vcc 700 mA Salida 2 2 x 5 Vcc 2 x 1 15 A M dulo de extensi n ri former Entrada 5 Vcc 3A 15 W Salida 1 1 x 3 5 Vcc 700 mA Salida 2 1 x 5 Vcc 1 x 1 15 A Clasificaci n Componente de aplicaci n ti...

Страница 41: ...as Esto es especialmente cierto cuando el ojo ha sido expuesto a la fotograf a retiniana La luz de este instrumento puede ser da ina El riesgo de da o ocular aumenta con la duraci n de la irradiaci n...

Страница 42: ...V 3 5 V 280 mA promedio de vida til 10 000 h 6 6 Esp culo nasal 6 6 1 Uso previsto El esp culo nasal descrito en estas instrucciones de funcionamiento ha sido fabricado para iluminar y por lo tanto ex...

Страница 43: ...o en el cabezal del instrumento para que la base de la l mpara encaje en la ranura del cabezal del instrumento 8 Instrucciones de conservaci n 8 1 Nota general La limpieza y desinfecci n de los dispos...

Страница 44: ...da para usarse en el entorno electromagn tico especificado a conti nuaci n El cliente o el usuario de la unidad de diagn stico ri former debe garantizar que se use en dicho entorno Gu a y declaraci n...

Страница 45: ...ubicaci n de la unidad de diagn stico ri former b En el rango de frecuencias de 150 kHz a 80 MHz las intensidades de campo deben ser inferiores a 3 V m Distancias de separaci n recomendadas entre equi...

Страница 46: ...LED 3 5 V LED dispositivo antirrobo EliteVue N de art culo 10512 Cabezal EliteVue simple LED 3 5 V N de art culo 10512 301 Cabezal EliteVue simple LED 3 5 V con dispositivo antirrobo para estaci n de...

Страница 47: ...1 Informazioni generali 4 2 Pulizia e disinfezione 5 Dati tecnici 6 Teste strumentali ri scope L Teste strumentali ri scope 6 1 Otoscopio ri scope L 6 2 Oftalmoscopi ri scope L 6 3 Retinoscopi a fessu...

Страница 48: ...spositivo pu essere utilizzato da medici e personale infermieristi co presso ospedali strutture sanitarie cliniche e ambulatori vietato l uso negli ambienti RM 1 4 5 Competenze formazione richieste ag...

Страница 49: ...ture sanitarie cliniche e ambulatori Attenzione Responsabilit dell utente responsabilit dell utente Prima di ogni utilizzo verificare l integrit e la comple tezza della Stazione diagnostica ri former...

Страница 50: ...universale e inserirlo Assicurarsi che entrambi i perni siano inseriti nel morsetto universale Fissare quindi il dispositivo ri former anestesia serrando la vite laterale 3 1 3 Fissaggio delle piastr...

Страница 51: ...resso 5 VDC 3A 15 W Uscita 1 1 x 3 5 V dc 700 mA Uscita 2 2 x 5 V dc 2 x 1 15 A Modulo di estensione ri former Ingresso 5 VDC 3A 15 W Uscita 1 1 x 3 5 Vdc 700 mA Uscita 2 1 x 5 Vdc 1 x 1 15 A Classifi...

Страница 52: ...oporzionale alla durata dell irraggiamento Il valore standard del rischio viene superato quando la durata dell irradiazione luminosa proveniente da questo strumento superiore ai 5 minuti Nonostante no...

Страница 53: ...a bat terie e ruotarlo in senso orario fino alla posizione finale Rimuovere la testa ruotando in senso antiorario Per due modalit operative Divaricazione rapida Spingere con il pollice la vite di rego...

Страница 54: ...ali e dei manici Pulire con disinfettante seguendo le istruzioni del produttore del disin fettante Devono essere utilizzati solo disinfettanti di comprovata effica cia in considerazione delle linee gu...

Страница 55: ...io ne diagnostica ri former devono assicurarsi che sia utilizzata in ambienti corrispondenti Test delle emissioni Conformit Ambiente elettromagnetico guida Emisiones de radiofrecuencia CISPR 11 Gruppo...

Страница 56: ...ssarie misure aggiuntive quali il riorientamento o il riposizionamento della Stazione diagnostica ri former b Oltre la gamma di frequenza da 150 kHz a 80 MHz le intensit di campo devono essere inferio...

Страница 57: ...3 5 V Art n n 10512 301 Testa EliteVue singola LED 3 5 V con dispositivo anti furto per stazione diagnostica a parete ri former Art n n 10513 Testa EliteVue singola XL 3 5 V Art n n 10513 301 Testa E...

Страница 58: ...3 1 4 4 1 4 5 1 4 6 1 5 1 6 2 2 1 2 2 3 3 1 3 2 3 3 3 4 ri former Mobile 3 5 ri former Anesthesia 4 4 1 4 2 5 6 ri scope L ri scope 6 1 ri scope L 6 2 ri scope L 6 3 6 4 6 5 6 6 6 7 6 8 6 9 6 10 7 L1...

Страница 59: ...59 1 3 IEC 60601 1 2 1 4 ri former 1 4 1 ri former C F 2002 96 EC Max 1 5 kg Mass 13 kg ri former big ben 1 4 2 1 4 3 1 4 4 1 4 5 1 4 6 1 5 ri former 2...

Страница 60: ...ri former ri former ri former ri former Ries ter 3650 ri former 1 3 5 100 240 3652 3652 ri former 2 3 5 100 240 3650 300 ri former 2 3 5 100 240 3652 300 ri former 2 3 5 100 240 ri former ri former 1...

Страница 61: ...2 3 3 3 3 1 HR MIN 3 3 2 3 3 3 100 3 3 4 rheotronic 3 4 ri former Mobile ri former 3 5 ri former Anesthesia ri former An esthesia ri former Anesthesia 3 1 3 3 1 4 a b c 3 1 5 b c I ON 0 OFF HR MIN B 3...

Страница 62: ...ormer 100 240 50 60 0 6 5 3 15 ri former 5 3 15 1 1 3 5 700 2 2 5 2 1 15 ri former 5 3 15 1 1 3 5 700 2 1 5 1 1 15 B 0 C 40 C 10 85 6 ri scope L ri scope ri scope L ri scope L b a b a 5 C 50 C 10 85 7...

Страница 63: ...3 6 1 3 360 6 1 4 180 3 6 1 5 6 1 6 XL 2 5 2 5 750 15 XL 3 5 3 5 720 15 LED 2 5 2 5 280 10 000 LED 3 5 3 5 280 10 000 6 2 ri scope L 6 2 1 Riester 24 5 DIN EN 62471 2 3 6 2 2 L1 L2 1 10 12 15 20 40 1...

Страница 64: ...ma 6 4 1 ri derma 6 4 2 6 4 3 6 4 4 2 0 10 6 4 5 ri derma XL 2 5 2 5 750 15 ri derma XL 3 5 3 5 690 15 ri derma LED 2 5 2 5 280 10 000 ri derma LED 3 5 3 5 280 10 000 24 6 5 6 5 1 6 5 2 6 5 3 XL 2 5 2...

Страница 65: ...5 280 10 000 6 8 6 8 1 Riester 6 8 2 6 9 6 9 1 Riester 6 9 2 6 9 3 360 2 5 6 9 4 6 9 5 HL 2 5 2 5 680 20 XL 3 5 3 5 700 20 6 10 6 10 1 Riester 6 10 2 6 10 3 2 5 360 6 10 4 6 10 5 HL 2 5 2 5 680 40 XL...

Страница 66: ...1 5 IEC 60601 1 30 12 ri former ri former ri former ri former ri former CISPR 11 1 ri former CISPR 11 B La Stazione diagnostica ri for mer idonea all uso in tutti gli ambienti compresi gli ambienti re...

Страница 67: ...2 0 IEC 61000 4 11 0 UT 0 5 1 0 UT 1 70 UT 25 30 50 60 0 UT 0 5 1 0 UT 1 70 UT 25 30 50 60 0 UT 250 300 50 60 0 UT 250 300 50 60 50 60 IEC 61000 4 8 30 A 50 30 A 50 UT ri former ri former IEC 60601 IE...

Страница 68: ...scope L 10537 10537 301 3 5 V XL 3 5 V XL 10575 10575 301 3 5 V LED 3 5 V LED ri scope L portalampadina a fibra ottica 10539 10539 301 3 5 V XL 3 5 V XL 10576 10576 301 3 5 V LED 3 5 V LED EliteVue 10...

Страница 69: ...69 12 2 B R1 shock proof 5 CE P Riester Rudolf Riester GmbH Dept Repairs RR Bruckstr 31 72417 Jungingen Germany...

Страница 70: ...70...

Страница 71: ...71...

Страница 72: ...417 Jungingen Deutschland Tel 49 7477 9270 0 Fax 49 7477 9270 70 info riester de www riester de 99249 Rev 0_2020 05 nderungen vorbehalten Subject to alterations Sous reserve de modifications Sujeto a...

Отзывы: