Wipe-disinfection as specified by the disinfectant manufacturer. Only disinfectants 
with proven efficacy should be used, taking into account the national requirements. 
After disinfection, wipe the instrument down with a moist cloth to remove possible 
disinfectant residues.

PLEASE NOTE!

 Never immerse the handles in liquids! Take care to ensure that no 

liquids get inside the casing! This item is not approved for automated reprocessing 
and sterilization. These procedures cause irreparable damage!

3. Otoscope and accessories
3.1. Purpose
Riester

 otoscopes described in these Instructions for Use have been produced for 

lighting and examination of the auditory canal, combined with a 

Riester

 ear specu-

lum.

3.2. Insertion and removal of ear speculum

Position the selected speculum on the chromium plated metal socket of the otosco-
pe. Turn speculum to the right until a resistance is felt. The size of the speculum is 
marked on the reverse.

3.3. Swivel lens for magnification

The swivel lens is fixed to the device and can be swivelled 360°.

3.4. Insertion of external instruments into the ear

If you wish to insert external instruments into the ear (e.g. tweezers), you have to rota-
te the swivel lens (approx. 3-fold magnification) located on the otoscope head by 180°.

3.5. Pneumatic test

In order to perform a pneumatic test (= examination of the ear drum), you will require 
a bulb which is not included in the normal scope of supply but may be ordered sepa-
rately (see Spare parts and accessories). Take metal connector which is not included 
in the normal scope of supply but may be ordered separately (see Spare parts and 
accessories) and insert in recess provided on the side of the otoscope head. Attach 
hose of bulb to connector. Carefully introduce the required air volume into the au-
ditory canal.

3.6. Replacement of lamp

e-

scope

®

 otoscope with direct illumination Remove the speculum socket by turning 

it to the left with your thumb and index finger until it stops. Pull the speculum so-
cket forward to remove it. Unscrew the bulb counterclockwise. Screw the new bulb in 
clockwise and reattach the speculum socket.

e-

scope

®

 

otoscope with fiber optics

Unscrew the instrument head from the battery handle. The LED/bulb is located in the 
lower part of the instrument head. Pull the bulb out of the instrument head using your 
thumb and index finger or a suitable tool. When replacing an LED with a bulb, the op-
tionally available adapter is additionally required; when replacing a bulb with an LED, 
the adapter must first be removed from the bulb unit. Firmly insert the new LED/bulb.

3.7 Instructions for care
General information

Cleaning and disinfection of the medical devices serves to protect the patient, the 
user and third parties and to preserve the value of the medical devices. Due to the 
product design and the materials used, no defined limit can be specified for the ma-
ximum number of reprocessing cycles that can be carried out. The life span of the 
medical devices is determined by their function and by gentle handling of the devices. 
Defective products must undergo the reprocessing procedure described before being 
returned for repair.

Cleaning and disinfection

The otoscope can be cleaned externally with a moist cloth until visually clean. Wi-
pe-disinfection as specified by the disinfectant manufacturer. Only disinfectants with 
proven efficacy should be used, taking into account the national requirements. After 
disinfection, wipe the instrument down with a moist cloth to remove possible disin-
fectant residues.

PLEASE NOTE!

 Never immerse the otoscope in liquids! Take care to ensure that no 

liquids get inside the casing! This item is not approved for automated reprocessing 
and sterilization. These procedures cause irreparable damage!

Sterilization
a) Reusable ear specula

Содержание e-scope e-xam

Страница 1: ...isung Diagnostische Instrumente Instructions Diagnostic Instruments Mode d emploi Instruments diagnostiques Instrucciones para el uso Instrumentos diagn sticos Istruzioni per I uso Strumenti diagnosti...

Страница 2: ...it Energie die Lampen sind in den entsprechen den Instrumentenk pfen enthalten Sie dienen ferner als Halter 2 2 Betriebsbereitschaft Einlegen und Herausnehmen von Batterien Drehen Sie den Instrumenten...

Страница 3: ...chf hren zu k nnen ben tigen Sie einen Ball der im normalen Lieferumfang nicht enthalten ist aber zus tzlich bestellt werden kann siehe Ersatzteile und Zubeh r Nehmen Sie den Metallkonnektor der im no...

Страница 4: ...malverwendbare Ohrtrichter 2 mm 100 St ck Art Nr 14061 532 500 St ck Art Nr 14062 532 1000 St ck Art Nr 14063 532 2 5 mm 100 St ck Art Nr 14061 531 500 St ck Art Nr 14062 531 1000 St ck Art Nr 14063 5...

Страница 5: ...wendet werden beim Wechsel von Gl hlampe auf LED muss dieser aus dem Lampe neinsatz entfernt werden Setzen Sie die neue LED Gl hlampe fest ein ACHTUNG Der Stift der Lampe muss in die F hrungsnut am A...

Страница 6: ...r der Entz ndung wenn das Ger t in Anwe senheit von brennbaren Gemischen von Arzneimittel mit Luft bzw mit Sauerstoff Lachgas oder An sthesiegasen betrieben wird Sicherheitsinformationen gem In ternat...

Страница 7: ...07...

Страница 8: ...08...

Страница 9: ...ration insertion and removal of batteries Turn off instrument head from handle in counter clockwise direction Insert two com mercial type AA Mignon alkaline batteries of 1 5 V IEC standard reference L...

Страница 10: ...al 3 6 Replacement of lamp e scope otoscope with direct illumination Remove the speculum socket by turning it to the left with your thumb and index nger until it stops Pull the speculum so cket forwar...

Страница 11: ...Packung 6 St ck No 10600 LED 3 7 V No 14041 Technical data of the lamp for e scope otoscope with direct illumination Vacuum 2 5 V 300 mA mean life span 15h XL 2 5 V 750 mA mean life span 16 5h for e...

Страница 12: ...e span 20000h Technical data of the e xam lamp XL 2 5V 750 mA mean life span 16 5h LED 2 5V 120 mA 5000 5500 Kelvin CRI 72 mean life span 20 000h 4 6 Instructions for care General information Cleaning...

Страница 13: ...equipment is subject to special precautionary measures with re gard to electromagnetic compatibility EMC Portable and mobile high frequency communication equipment can in uence medical electrical equi...

Страница 14: ...14...

Страница 15: ...servit en outre de support 2 2 tat de marche mise en place et retrait des piles D vissez du manche la t te d instrument dans le sens inverse des aiguilles d une montre Placez 2 piles alcalines du comm...

Страница 16: ...oir Pi ces de rechange et accessoires Prendrez le connecteur m tallique et chez le lat ralement sur la t te d otoscope dans l videment pr vu cet effet Le tuyau de la poire est ch sur le connecteur Vou...

Страница 17: ...3 7 V R f 14041 Caract ristiques techniques de la lampe pour otoscope e scope avec clairage direct Vide 2 5 V 300 mA dur e de vie moy 15 h XL 2 5 V 750 mA dur e de vie moy 16 5 h pour otoscope e scop...

Страница 18: ...ite 4 5 Caract ristique techniques de la lampe d ophthalmoscope XL 2 5 V 750 mA dur e de vie moy 16 5 h LED 3 7 V 38 mA dur e de vie moy 20 000 h Caract ristique technique de la lampe e xam XL 2 5V 75...

Страница 19: ...ationale europ enne en vigueur 7 Compatibilit lectromagn tique Les appareils lectriques m dicaux doivent faire l objet de mesures de pr caution sp ciales concernant la compatibilit lectromagn tique CE...

Страница 20: ...20...

Страница 21: ...diente formaci n t cnica y profesional Para que podamos avalar un funcionamiento perfecto y able de nuestros instrumentos es imprescin dible utilizarlos exclusivamente con accesorios de la casa Rieste...

Страница 22: ...r de la carcasa El art culo no est aprobado para la esterilizaci n ni el reacondicionamiento autom ticos Podr an producirse da os irreparables 3 Otoscopio y accesorios 3 1 Aplicaci n apropiada Los oto...

Страница 23: ...tante Se deben utilizar exclusivamente productos de e cacia probada y teniendo en cuenta los requisitos nacionales Tras la desinfec ci n limpie el instrumento con un pa o h medo para eliminar los posi...

Страница 24: ...naci n de jacionescentrales o exc ntricas Filtros Funci n Filtro exento de rojo ltro verde refuerzo de contraste para la evaluaci n de alteraciones min sculas de los vasos sangu neos por ejemplo hemor...

Страница 25: ...por favor dir jase a nuestra empresa o a un representante reconocido por Riester le asistiremos gustosamente 6 Observaci n Temperatura ambiental 0 hasta 40 C Humedad Relativa 30 hasta 70 no conensado...

Страница 26: ...26...

Страница 27: ...27...

Страница 28: ...o dal manico svitandola in senso anti orario Inserire nel manico 2 batterie alcaline comuni del tipo AA stilo da 1 5 V denominazione IEC LR6 in modo tale che i poli positivi siano rivolti verso la par...

Страница 29: ...evista sul lato della testa dell otoscopio Il tubo della pompetta sferica viene inserito sul connettore Ora possibile immettere con attenzione nel canale auricolare solo la quantit d aria necessaria 3...

Страница 30: ...per otoscopio e scope con illuminazione diretta A vuoto 2 5 V 300 mA durata media 15h XL 2 5 V 750 mA durata media 16 5h per e scope otoscopio a bra ottica XL 2 5 V 750 mA durata media 15h LED 3 7 V 5...

Страница 31: ...ata media 16 5h LED 3 7 V 38 mA durata media 20000h Dati tecnici della lampadina e xam XL 2 5V 750 mA durata media 16 5h LED 2 5V 120 mA 5000 5500 Kelvin CRI 72 durata media 20 000h 4 6 Avvertenze per...

Страница 32: ...ggetti a particolari misure precauzionali per quanto riguarda la compatibilit elettromagnetica CEM I dispositivi mobili e portatili di comunicazione ad alta frequenza possono in uire sul funzionamento...

Страница 33: ...33...

Страница 34: ...34 RUSSIAN 1 2 3 3 4 e xam 5 Mantenimiento 5 6 7 1 Riester 93 42 E C H E Riester Riester Riester Riester B 2 2 1 H P Riester 2 2 alkaline AA 1 5 IEC LR6 B E B B...

Страница 35: ...35 2 3 0 2 4 B 2 5 3 3 1 H Riester Riester 3 2 P P 3 3 360 3 4 B 3 180 3 5 B B B 3 6 e scope e scope...

Страница 36: ...5 10 0795 2 100 14061 532 500 14062 532 1000 14063 532 2 5 100 14061 531 500 14062 531 1000 14063 531 3 100 14061 533 500 14062 533 1000 14063 533 4 100 14061 534 500 14062 534 1000 14063 534 5 100 14...

Страница 37: ...300 15 XL 2 5 750 16 5 e scope F O XL 2 5 750 15 3 7 52 20000 10960 2 10961 7 4 e xam 4 1 H Riester e xam EN 62471 2008 4 2 D 1 2 3 4 6 8 10 15 20 D 1 2 3 4 6 8 10 15 20 4 3 M 0 A e xam e xam 4 4 e s...

Страница 38: ...5h LED 2 5B 120 mA 5000 5500 Kelvin CRI 72 20 000h 4 6 e xam e xam 4 7 e scope XL 2 5 6 10605 3 7 14051 e xam XL 2 5B 6 N 11178 LED 2 5B N 12320 5 T B Ries ter Riester B B B 6 0 40 C 30 70 H 10 55 10...

Страница 39: ...39...

Страница 40: ...40...

Страница 41: ...he strictest quality standards and has undergone a thorough nal quality check before leaving our factory We are therefore pleased to be able to provide a warranty of 2 years from the date of purchase...

Страница 42: ...e la fecha de compra por todos los fallos debidos demostrablemente a fallos de material o de fabricaci n La garant a quedar anulada en caso de utilizaci n indebida Durante el plazo de vigencia de la g...

Страница 43: ...tore in tutte le sue parti e debita mente provvista di timbro Fare attenzione che i diritti di garanzia vanno fatti valere entro il periodo di garanzia Eventuali veri che o riparazioni successivamente...

Страница 44: ...44...

Страница 45: ...45...

Страница 46: ...46...

Страница 47: ...47...

Страница 48: ...72417Jungingen Germany Tel 49 7477 9270 0 Fax 49 7477 9270 70 E Mail info riester de www riester de 99231 Rev B 2015 03 nderungen vorbehalten Subject to alterations Sous r serve de modi cations Sujet...

Отзывы: