3.3 SERVICE
TOOLS
S12 Temperature
Calibration
The correction of the temperature regulation is possible
by input of the temperature value which is measured in
the cuvette with an external measuring system
(Thermistor and DMM).
This function is available only when the temperature
regulation is stabilized. In an unstable condition an
error message is shown on the display.
Enter the actual temperature value always four-digit in
° C e.g.
[3]
[7]
[.]
[2] [0]
measured by an external
measuring system.
Confirm with [E].
The system calculates and stores the new setting for
the selected temperature level and starts working with it
at once.
T E M P . C A L I B R A T I O N
T E M P . A C T U A L :
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
S13 ADC
Correction
The current ADC correction of the device can be read
and set.
The range of this value is 0 up to 255. The default
value is 150.
The input has to be confirmed with [E].
A D C C O R R E C T I O N 1 5 0
1 2 3 4 5 6 7 8
E S C 9 0 . + - E
S14
ADC Counts (Optic)
Indicated is the current value of the optical analogue-
digital converter. The value is proportional to the light-
current depending on the selected amplification level
and the boost setting.
To a key actuation the system reacts possibly only after
three seconds.
The functions V0 to V7 set the amplification level. The
function BO increases (boost ON) or reduces (boost
OFF) the time slot. The functions F0 to F9 place the
filter wheel into the positions 0 to 9. The position 0
corresponds to the filter wheel position at the dark level
adjustment.
Stop function by [ESC].
A D C C O U N T S ( O P T I C )
B O O S T O F F
V 0 F 1 3 4 0 n m 1 8 8 1 0
V 0 V 1 V 2 V 3 V 4 V 5 V 6 V 7 B 0 .
F 0 F 1 F 2 F 3 F 4 F 5 F 6 F 7 F 8 F 9
E S C
RIELE BERLIN
Photometer 5010 V5+
3.3.10
11.03.2009
Содержание Photometer 5010 V5+
Страница 1: ...Photometer 5010 V5 Version 5 x SN 7000 Service Manual ...
Страница 2: ...3 1 BLOCK DIAGRAM up to SN 8001 RIELE BERLIN Photometer 5010 V5 3 1 1 29 08 2007 ...
Страница 3: ...3 1 BLOCK DIAGRAM from SN 8002 RIELE BERLIN Photometer 5010 V5 3 1 2 29 08 2007 ...
Страница 78: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 14 04 08 2008 ...
Страница 79: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 15 04 08 2008 ...
Страница 80: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 16 04 08 2008 ...
Страница 81: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 17 04 08 2008 ...
Страница 82: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 18 04 08 2008 ...
Страница 83: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 33 R RIELE BERLIN Photometer 5010 V5 4 40 19 04 08 2008 ...
Страница 85: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 21 04 08 2008 ...
Страница 86: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 22 04 08 2008 ...
Страница 87: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 23 04 08 2008 ...
Страница 88: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 24 04 08 2008 ...
Страница 89: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 25 04 08 2008 ...
Страница 90: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 26 04 08 2008 ...