4.40 PCB RR-120 MOTHERBOARD REF
5010-120
Installation instruction
In case of changing the mother board following items have to be done after switching on the device:
1. Check the contrast of LCD
(Service Manual chapter 4.42 – Contrast of touchscreen).
2. Check the voltage 12 V
DC
of halogen lamp. The voltage was pre-adjusted at the factory
(Service Manual chapter 4.40 – Voltage 12 V
DC
of halogen lamp).
3. Check vo 12 V
DC
, - 12 V
DC
, + 5 V
DC
and + 3 V
DC
(battery power)
(Service Manual chapter 4.40 – Voltages).
4. Perform the adjustment of amplifier stage
(Service Manual chapter 4.40 – Adjustment of amplifier stage).
5. The display of Photometer
5010 V5+
shows:
M A I N M E N U
M E A S U R E W I T H P R O G R . M E T H O D S
M E A S U R E W I T H B A S I C M E T H O D S
M E T H O D S N E W / C H A N G E / C O P Y
U T I L I T I E S L A M P L F
3 7 . 0 5 C 0 4 . 0 1 . 0 6 1 1 : 3 9
6. The serial number can be set by applying the service tool
S5
(Service Manual chapter 3.3 – S5 - Serial no.).
7. The assignment of filter positions has to be done if additional filters has been built in
(e.g. 670 nm). Apply the service tool
S7
(Service Manual chapter 3.3 – S7 – Set wavelength).
8. Do a setting of the temperature values by performing the calibration procedure of
Photometer
5010 V5+.
(Service Manual chapter 9.1 - Subject: Cuvette temperature)
(Service Manual chapter 4.40 - Thermal adjustment)
9. Do a setting of “attribute of PCB layout“ by applying the service tool
S6
:
Software applications may be dependent on this attribute
(Service Manual chapter 3.3 – S6 - PCB layout).
10. Delete allocated memory by applying the service tool
S2
:
(Service Manual chapter 3.3 – S2.1 – Delete free methods)
(Service Manual chapter 3.3 – S2.2 – Multi-standard memory)
(Service Manual chapter 3.3 – S2.3 – Delete logger memory)
(Service Manual chapter 3.3 – S2.4 – Delete QC memory)
11. Define the laboratory name by applying the utility
U11
:
The name of the laboratory can be stored permanently
(OM chapter 7.2.11 – Laboratory name).
12. Define the user names by applying the utility
U12
:
The names of maximum five users can be stored permanently
(OM chapter 7.2.12 – User name).
13. Do a dark level adjustment by applying utility
U1
in combination with flow-through system.
(OM chapter 7.2.1 - Dark level adjustment).
The utility is working specifically to the current adaptor. An adjustment of the standard cuvette adaptor
should be done later on.
14. Perform a media test by applying utility
U4.4
(OM chapter 7.2.4.2 – Media test).
RIELE BERLIN
Photometer 5010 V5+
4.40.12
17.03.2009
Содержание Photometer 5010 V5+
Страница 1: ...Photometer 5010 V5 Version 5 x SN 7000 Service Manual ...
Страница 2: ...3 1 BLOCK DIAGRAM up to SN 8001 RIELE BERLIN Photometer 5010 V5 3 1 1 29 08 2007 ...
Страница 3: ...3 1 BLOCK DIAGRAM from SN 8002 RIELE BERLIN Photometer 5010 V5 3 1 2 29 08 2007 ...
Страница 78: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 14 04 08 2008 ...
Страница 79: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 15 04 08 2008 ...
Страница 80: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 16 04 08 2008 ...
Страница 81: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 17 04 08 2008 ...
Страница 82: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 18 04 08 2008 ...
Страница 83: ...4 40 PCB RR 120 MOTHERBOARD Layout c REF 5010 120 33 R RIELE BERLIN Photometer 5010 V5 4 40 19 04 08 2008 ...
Страница 85: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 21 04 08 2008 ...
Страница 86: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 22 04 08 2008 ...
Страница 87: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 23 04 08 2008 ...
Страница 88: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 24 04 08 2008 ...
Страница 89: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 25 04 08 2008 ...
Страница 90: ...4 40 PCB RR 120 MOTHERBOARD Layout d REF 5010 120 RIELE BERLIN Photometer 5010 V5 4 40 26 04 08 2008 ...