Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
21
3.2.6 Program Menu
Figure 3-8. Program Menu
PROGRM Menu
Parameter
Choices
Description
Level 2 submenus
PWRUPM
GO
DELAY
Power up mode. In GO mode, the indicator goes into operation immediately after a brief power up display test.
In DELAY mode, the indicator performs a power up display test, then enters a 30-second warm up period. If no motion
is detected during the warm up period, the indicator becomes operational when the warm up period ends; if motion is
detected, the delay timer is reset and the warm up period repeated.
REGULA
NTEP
OIML
CANADA
NONE
Regulatory mode. Specifies the regulatory agency having jurisdiction over the scale site.
OIML, NTEP, and CANADA modes allow a tare to be acquired at any weight greater than zero. NONE allows tares to
be acquired at any weight value.
OIML, NTEP, and CANADA modes allow a tare to be cleared only if the gross weight is at no load. NONE allows tares
to be cleared at any weight value.
NTEP and OIML modes allow a new tare to be acquired even if a tare is already present. In CANADA mode, the pre
-
vious tare must be cleared before a new tare can be acquired.
NONE, NTEP and CANADA modes allow the scale to be zeroed in either gross or net mode as long as the current
weight is within the specified ZRANGE. In OIML mode, the scale must be in gross mode before it can be zeroed;
pressing the ZERO key in net mode clears the tare.
UID
1
Specifies the unit identification number. Value specified can be any numeric value, up to six digits
DATE
DATFMT
DATSEP
Allows selection of date format and date separator. See Level three parameter for descriptions
TIME
TIMFMT
TIMSEP
Allows selection of time format and separator. See level three parameter for descriptions
CONSNU
000000
number
Consecutive numbering. Allows sequential numbering for print operations. The consecutive number value is incre
-
mented following each print operation.
The initial value of this parameter is set to the start up value specified on the CONSTU parameter. Changing either
CONSTU or CONSNU immediately resets the consecutive number used for printing.
CONSTU
000000
number
Consecutive number start up value. Specifies the initial consecutive number (CONSNU) value used when the indica
-
tor is powered on
Level 3 submenus
DATFMT
MMDDYY
DDMMYY
YYMMDD
Specifies the format used to display the date
Table 3-8. Program Menu Parameters
XXXXXXX
XXXXXXX
PWRUPM
GO
OIML
REGULA
000000
number
CONSNU
DELAY
SETPNT
PROGRM
SERIAL
CALIBR
CONFIG FORMAT
000000
number
CONSTU
NTEP
CANADA
1
UID
VERS
DA TFMT
MMDDYY
DA TE
DDMMYY
YYMMDD
YYDDMM
DA TSEP
SLASH
DASH
SEMI
TIMFMT
12HOUR
TIME
24HOUR
TIMSEP
COLON
COMA
NONE
DEFCAL