Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
15
3.2.2 Configuration Menu
Figure 3-4. Configuration Menu
CONFIG Menu
Parameter
Choices
Description
Level 2 submenus
GRADS
1000
number
Graduations. Specifies the number of full scale graduations. The value entered must be in the range 1–100 000 and
should be consistent with legal requirements and environmental limits on system resolution.
To calculate GRADS, use the formula, GRADS =
Capacity
/
Display Divisions
.
Display divisions for primary and secondary units are specified on the FORMAT menu.
ZTRKBN
0
number
Automatically zeros the scale when within the range specified, as long as the input is within the ZRANGE and scale
is at standstill. Specify the zero tracking band in ± display divisions. Maximum legal value varies depending on local
regulations.
Note: For scales using linear calibration, do not set the zero tracking band to a value greater than that spec
-
ified for the first linearization point.
ZRANGE
1.900000
number
Selects the range within which the scale can be zeroed. The 1.900000 default value is ± 1.9% around the calibrated
zero point, for a total range of 3.8%. Indicator must be at standstill to zero the scale. Use the default value for legal-
for-trade applications.
Table 3-4. Configuration Menu Parameters
XXXXXXX
XXXXXXX
GRADS
1000
ZTRKBN
ZRANGE
1.900000
MOTBAN
FS+2%
FS+9D
FS+1D
FS
OVRLOA
1
4
2
8
DIGFL1
32
16
64
1
4
2
8
32
16
64
1
4
2
8
32
16
64
number
DIGFL2
DIGFL3
BOTH
PBTARE
NOT ARE
KEYED
T AREFN
SETPNT
PROGRM
SERIAL
CALIBR
CONFIG FORMAT
8OUT
32OUT
16OUT
64OUT
DFSENS
DFTHRH
2OUT
128OUT
4OUT
2DD
NONE
5DD
20DD
10DD
50DD
200DD
100DD
250DD
15HZ
60HZ
30HZ
7.5HZ
SMPRAT
VERS
DEFCAL
0
1
number
number
number