680HE Synergy Indicator
36
Visit our website
4.4.8.2 If KIND = ACCUM, DELAY, WAITSS, AUTOJG, TOD
Figure 4-19. Setpoints – Parameter Group B
4.4.8.3 If KIND = PAUSE, COUNTR, COZ, INMOTN, INRNGE, BATCHPR, TIMER, CONCUR
Figure 4-20. Setpoints – Parameter Group C
Parameter
Description
value
Setpoint Value – For weight-based setpoints:
Enter value: 0.0–99999.0,
0.0
(default)
For time-based setpoints:
Enter value: 0.0–65535.0,
0.0
(default)
For COUNTR setpoints:
Enter value: 0.0–65535.0,
0.0
(default)
trip
Trip – Specifies whether the setpoint is satisfied when the weight is higher or lower than the setpoint value, within a band established
around the value, or outside of the band; in a batch sequence with TRIP=HIGHER, the associated digital output is active until the setpoint
value is reached or exceeded; with TRIP=LOWER, the output is active until the weight goes below the setpoint value;
Settings:
HIGHER
(default), LOWER, INBAND, OUTBND
bndval
Band Value – For setpoints with TRIP=INBAND or OUTBND, specifies a weight equal to half the band width; The band established
around the setpoint value is VALUE ±BNDVAL;
Enter value: 0.0–99999.0,
0.0
(default)
hystrs
Hysteresis – Specifies a band around the setpoint value which must be exceeded before the setpoint, once off, can trip
on again;
Enter value: 0.0–99999.0,
0.0
(default)
preact
Preact Type – Allows the digital output associated with a setpoint to shut off before the setpoint is satisfied to allow for
material in suspension;
Settings:
OFF
(default)
ON – Adjusts setpoint trip value up or down (depending on the TRIP parameter setting) from setpoint value using a fixed value
specified on the PREVAL parameter
LEARN – Can be used to automatically adjust the PREACT value after each batch; it compares the actual weight at standstill to the
target setpoint value and adjusts the PREVAL by the PREADJ value times the difference after each batch
Table 4-17. Kind Parameter Descriptions
setpt1-8
sptcfg
kind
value
trip
bndval
time
durtn
alarm
clracc
clrtar
pshacc
pshprt
pshtar
dslot
digout
sense
batch
branch
access
enable
alias
If KIND = ACCUM
or DELAY
If KIND = TOD
If KIND = ACCUM
or TOD
If BATCH = ON and
KIND = ACCUM
If KIND = ACCUM
If KIND = ACCUM
If TRIP = INBAND
or OUTBND
If KIND = ACCUM,
DELAY or WAITSS
setpt1-8
sptcfg
kind
value
start
end
dslot
digout
sense
branch
access
alias
If KIND = COUNTR,
TIMER or CONCUR
If KIND = COUNTR
If KIND = TIMER or CONCUR