Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
23
4.4 Setup Menu
Figure 4-4. Setup Menu
4.4.1 Setup – Configuration Menu
Figure 4-5. Setup – Configuration Menu
Menu
Description
Config
Configuration – See
for menu structure and parameter descriptions of the Configuration menu
Format
Format – See
for menu structure and parameter descriptions of the Format menu
Calibr
Calibration – See
for menu structure and parameter descriptions of the Calibration menu
Comm
Communication – See
for menu structure and parameter descriptions of the Communication menu
Progrm
for menu structure and parameter descriptions of the Program menu
Pformt
Print Format – See
for menu structure and parameter descriptions of the Print Format menu
sformt
Stream Format – See
for menu structure and parameter descriptions of the Stream Format menu
Setpnt
Setpoints – See
for menu structure and parameter descriptions of the Setpoint menu
Digio
for menu structure and parameter descriptions of the Digital I/O menu
Algout
for menu structure and parameter descriptions of the Analog Output menu
deflt
Default Configuration – See
for instructions to reset the configuration settings
Table 4-3. Setup Menu Descriptions
Menu
Description
Capcty
Capacity – Maximum rated capacity of the scale;
Enter value: 0.00001–99999.0,
10000.0
(default)
Ztrkbn
Zero Track Band – Automatically zeros the scale when within the range specified, as long as the input is within the
zrange
and
scale standstill; When weight is within the zero band, the center of zero annunciator displays; Max legal value depends on local
regulations; Specify the zero tracking band in ± display divisions;
Enter value: 0.0–100.0,
0.0
(default)
Zrange
Zero Range – The total amount the scale can be zeroed; Zero range represents a percentage of capacity; The default value of
1.9 represents ±1.9% around the calibrated zero point, for a total range of 3.8%; A value of 0.0 prevents zeroing; Maximum legal
value depends on local regulations;
Enter value: 0.0–100.0,
1.9
(default)
Intzro
Initial Zero Range – When the indicator is turned on and the weight value is between the ± percent range specified in Calibrated
Zero, the indicator automatically zeros off the weight;
Enter value: 0
.
0–100.0,
0.0
(default)
Motban
Motion Band – Sets the level, in display divisions, at which scale motion is detected; If motion is not detected for the time defined
by
ss time
, the standstill symbol lights; Some operations, including print, tare, and zero, require the scale to be at standstill;
Maximum legal value varies depending on local regulations; If this parameter is set to 0, the standstill annunciator is always lit
and operations requiring standstill are performed regardless of scale motion; If 0 is selected,
ztrkbnd
must also be set to 0;
Enter value: 0–100,
1
(default)
Table 4-4. Setup – Configuration Menu Descriptions
Setup
Config
Format
Calibr
Comm
Progrm
sFormt
Setpnt
DigIO
AlgOut
deflt
pformt
Setup
Config
Capcty
fltchn
digfl1
digfl2
digfl3
dfsens
dfthrh
adsens
adthrh
dmpvl
Ztrkbn
Zrange
Intzro
motban
ovrloa
smprat
rtltrp
tarefn
sstime
sense