Ishida OMNi-4000 Error Codes and Troubleshooting
15.8 ERROR CODE (1600 - )
No.
Message
Symptom or Phenomenon
Response or Check
1600 "A MEMORY SETUP OF THE SCALE
IS UNUSUAL"
Scale (A/D, P-930) memory settings are abnormal. Recalibrate the scale.
If the error cannot be resolved or recurs later replace the
A/D board P-930.
1601 "SPAN ADJUSTMENT OF THE SCALE
INCOMPLETE"
Scale (A/D, P-930) span adjustment incomplete.
NOTE: Ignore sub-error numbers, all indicate the
same problem.
Recalibrate the scale.
If the error cannot be resolved or recurs later replace the
A/D board P-930.
1602 "INITIALIZATION THE SCALE IS NOT
COMPLETED"
Scale (A/D, P-930) initialization has not been
completed.
Recalibrate the scale.
If the error cannot be resolved or recurs later replace the
A/D board P-930.
1618 "CHECK THE SCALE"
Check that nothing is in touch with weighing
conveyor.
(Air currents, vibration)
Remove the weighing conveyor to check if anything is touching the
weighing conveyor and remove foreign objects if any.
Also check for air currents or vibration.
1619 "TAKE THE ARTICLE ON THE SCALE
PLATE"
An item remains on the scale.
1. Remove item from the weighing platter.
2. Check for mechanical obstructions at the scale and conveyor.
The scale zero calibration is out of range.
1. Recalibrate the scale.
2. Replace the load cell.
3. Replace the A/D board.
1621 "THERE IS NO RESPONSE FROM
THE SCALE"
There is no response from the scale (A/D, P-930).
Restore power after checking scale connection
01: Invalid text
02: Initializing
03: Out of range
04: Unstable
05: Unable to change between A/D
06: Calibration adjustment is not complete
07: Unable to send (Scale debug response)
08-16: WPL
17: Upper/lower limit error
18: Check scale
19: Remove item from the weighing platter
20: Power failure detected
21: No response from the scale board
(connection error)
22: Communication error (connection error)
23: Communication error (connection error)
Check harness, connector
Replace P-930(XJ1,4,5) - X292 - X291 - P-910(XJ8)
1622 "COMMUNICATION IMPOSSIBLE
WITH THE SCALE"
Cannot communicate with the scale (A/D, P-930)
Restore power after checking scale connection.
Check harness, connector
Replace P-930 (XJ1,4,5)-X292-X291-P-910(XJ8)
Rev. Feb 2015
33/39