Installation
©
Rice Lake Weighing Systems
●
All Rights Reserved
3
2.0
Installation
The
IQ plus 210
digital weight indicator is designed for easy setup and installation. All models are pre-configured and weight
calibrated before shipment, with the load cell connected to the indicator. For simple, standalone scale applications, hardware
installation consists of attaching the indicator, mounting assembly, and bench scale, then leveling the scale.
The IQ plus 2100 is weight calibrated before shipment from the factory, but recalibration at the installation site is
strongly recommended. Calibration by a certified scale technician is required for Legal-for-Trade applications.
2.1 Unpacking and Assembly
Immediately after unpacking, visually inspect the
IQ plus 210
to ensure all components are included and undamaged. The
shipping carton should contain the indicator, the indicator mount (tilt stand) and a parts kit. If any parts were damaged in
shipment, notify Rice Lake Weighing Systems and the shipper immediately.
The parts kit contains the items listed below:
• Capacity label (PN 42350)
• One grounding clamp (PN 53075), external tooth lock washer, (PN 15133), and kep nut (PN 14626) for digital input or
serial cable shield grounding against the enclosure.
• Two wing knobs with nylon washers for attaching the indicator to the mounting assembly, four rubber tilt stand feet (for tilt
stand option) or bench scale feet, jam nuts, lock washers, and cap screws (for column and attachment bracket options).
To install, attach the mounting assembly to scale base, then assemble the indicator to the mounting assembly. Place the bench
scale in the desired location then lift off the scale platter and locate the bubble level. Adjust the corner feet until the scale is level
and all feet are in contact with the support surface. Tighten the jam nuts against the scale to lock-in the level adjustment.
Plug power cord into power outlet to begin weighing.
The supply cord serves as the power disconnect for the IQ plus 2100. The power outlet supplying the indicator
must be installed near the unit and be easily accessible.
2.2 Factory Setup
The
IQ plus 210
is pre-configured and weight calibrated at the factory before shipment. Table 2-1 lists the default graduations
and count-by configuration for each scale capacity.
If scale configuration must be changed, see
for detailed configuration information. See
Scale Capacity
Count by
Configuration Parameters
DECPNT
DSPDIV
5 lb
0.001 lb
888.888
1D
10 lb
0.002 lb
888.888
2D
25 lb
0.005 lb
888.888
5D
50 lb
0.01 lb
8888.88
1D
100 lb
0.02 lb
8888.88
2D
2.5 Kg
0.0005 Kg
88.8888
5D
5 Kg
0.001 Kg
888.888
1D
10 Kg
0.002 Kg
888.888
2D
25 Kg
0.005 Kg
888.888
5D
50 Kg
0.01 Kg
8888.88
1D
All capacities are pre-configured for 5000 graduations (GRADS
parameter on CONFIG menu set to 5000).
Table 2-1. Default Scale Configurations