IQ plus 210 Installation Manual
18
Visit our website
3.2.5 Digital Input Menu
Figure 3-10. Digital Input Menu
3.2.6 Default Menu
The DEFLT menu is used to reset indicator configuration to the default values. There are no parameters associated with the
DEFLT menu. From the DEFLT menu, press
ENTER
(
ZERO
) to move down to the RESET parameter. Press
ENTER
again to reset
the indicator to the default configuration values or press
RIGHT
(
UNITS
) to cancel the indicator reset and return to the menu level
above.
NOTE:
All load cell calibration settings are lost when the RESET function is performed.
Figure 3-11. Default Menu
3.2.7 Version Menu
The VERS menu is used to check the software version installed in the indicator. There are no parameters associated with the
VERS menu: when selected, the indicator displays the installed software version number.
Figure 3-12. Version Menu
DIG IN Menu
Parameter
Choices
Description
Level 2 submenus
DIGIN1
DIGIN2
OFF
ZERO
UNITS
Specifies the function activated by digital inputs 1 and 2.
Table 3-6. Digital Input Menu Parameters