Rice Lake IDIMENSION 200 Скачать руководство пользователя страница 46

Administrator’s Section

43

Item Tracking

Tracking Configuration String

A string conveying configuration parameters for the internal calibration target finder module.

Default Values: -hcol 1.0 -colsens 0.3 -softThresh 1 -interpMethod 1 -blobSeg0 -fftSizes 512

Locking

Valid Values

Description

-hcol

Heavy color [0,1] used to threshold scene image

-colSens

Color sensitivity [0,1] used to threshold scene image

-softThresh

0: hard threshold; 1: soft threshold

-interpMethod

0: nearest neighbor; 1: bilinear; 2:quadratic

-blobSeg

1: use blob segmentation; 0: don’t use blob segmentation

-fftSizes

Size for all internal images and FFTs

Table 5-12. Tracking Configuration String

Parameters

Description

Values

Depth Tolerance (mm)

A tolerance value (in millimeters) used during the segmentation process of 
the depth data provided by the sensor. Adjacent depth values that are 
within this tolerance will be deemed to belong to the same object.

Default Value: 10 mm
Valid Values: 5 mm-15 mm

RGB Diff Threshold (%)

A pixel intensity threshold relative to the full image dynamic used during 
the segmentation process of the RGB data provided by the sensor. 

Default Value: 30%
Valid Values: 0-100%

Apply Depth Mask

Determines whether or not the system should account for vertical protru-
sions on the platform. These are very rare cases when the platform may 
have a backstop or a lip. 

Default Value: Off
Valid Values: On/Off

Disk Finder Radius (mm) The radius (in millimeters) of the Disk Finder disk.

Default Value: 25

Disk Finder Threshold

Score threshold value above which the disk finder will consider a match to 
be successful.

Default Value: 85
Valid Values: 0-100

Edge Threshold (%)

Gradient threshold value used during scene edge detection. The local 
image gradient must be above that value in order for the local feature to be 
considered as an edge.

Default Value: 30
Valid Values: 0-100

Table 5-13. Item Tracking

Parameters

Description

Values

Track Deviation
Threshold (mm)

A threshold value (in millimeters) below which the deviation of the object’s 
three dimensions should remain in order for the system to settle and lock 
onto the object. 

Default Value: 10 mm
Valid Values: 5 - 30 mm

Dim Rect Score
Threshold (%)

The rectangle score of an object measures how rectangular its outline 
appears to be. This parameter is a rectangle score threshold value below 
which any returned dimension will be ignored. 

Default Value: 30%
Valid Values: 0-100%

Cuboid Score (%)

The cuboid score of an object measures how box-like it appears to be. 
This parameter is a cuboid score threshold value below which the object 
will be deemed irregular.

Default Value: 80%
Valid Values: 0-100%

Table 5-14. Locking

Содержание IDIMENSION 200

Страница 1: ...ustin Ft Worth NicolScales com 800 225 8181 Contact Us Nicol Scales Measurement is an ISO Accredited Calibration Company that has provided calibration repair and sales of all types of weighing and mea...

Страница 2: ...iDimension 300 Industrial Installation Manual PN 169040...

Страница 3: ......

Страница 4: ...t manuals Proper care is your responsibility Failure to heed may result in serious injury or death Electric shock hazard For pluggable equipment the socket outlet must be installed near the equipment...

Страница 5: ...2 iDimension 300 1 2 iDimension 300 Parts Descriptions Figure 1 1 iDimension 300 Parts Industrial Scanning Head Optional Stand...

Страница 6: ...0 AC 3 Standard USB Ports Type A 1 USB Port Type B Host Indicator Light USB Indicator Light iDimension 300 is functioning correctly when all three lights are on and green 1 4 Underside Panel Ethernet...

Страница 7: ...try s plug requirements a different plug may need to be supplied If using a separate cable for DC OUT these cables must be less than three meters in length Calibration object Ethernet cable Not Suppli...

Страница 8: ...verify that you can communicate with iDimension 300 200 from a computer Use the ping command to confirm connectivity ping 169 254 1 1 If the ping command does not show it responding this may be due t...

Страница 9: ...cting Restart from any screen in the QubeVu Manager will give the user the option to either restart or reboot Figure 2 4 Restart Reboot Prompt Running This device is running correctly Starting iDimens...

Страница 10: ...e including the firmware version number the firmware CRC the serial number and the certificate number To exit from this screen click on the x in the upper right section of the information screen Figur...

Страница 11: ...n Tools to log in Figure 2 6 Login 2 Select Admin Tools 3 Log in with a username and password The default username and password are listed below Username admin Pasword password The username and passwo...

Страница 12: ...network Figure 2 9 Setup Menu Select Network from the setup menu There are two tabs Network Settings defines iDimension 300 as a network device in the enterprise network Network Security defines the s...

Страница 13: ...this device must be defined This name will be used to access the device from the Manager Tools in the future A hostname can be up to 15 characters For example http hostname IP Address If DHCP was chec...

Страница 14: ...Tab 1 Select the Date Time tab from the Setup General Settings screen iDimension 300 s current date and time are displayed Figure 2 13 Current Date Displays 2 Click on the Date icon to change the dat...

Страница 15: ...12 iDimension 300 6 Click Done to save the settings...

Страница 16: ...object directly onto the base plate Center the calibration object using the base plate marks Figure 3 1 Base Plate Marks 3 1 1 Define the Scale If not using a scale proceed to Section 3 1 2 on page 15...

Страница 17: ...ollowing format VENDOR ID PRODUCT ID Scale Type Description Auto iDimension 300 will attempt to identify the scale and if identified will use that scale None There is no scale attached to the iDimensi...

Страница 18: ...min Tools menu Figure 3 4 iDimension Admin Tools Menu 2 Select Camera Calibration from the Calibration Menu Figure 3 5 Camera Calibration 3 If iDimension 300 is not already set to Configuring select E...

Страница 19: ...llow diagonal lines If the marks cannot be centered try to adjust the physical position of the head by adjusting the top section of the head and or adjusting the screws on the rear of the head 6 Hold...

Страница 20: ...nd the Calibration Object 11 Press Calibrate Figure 3 10 Camera Calibration 12 All three tabs must show the green check mark for the calibration to be successful Review each tab to ensure that the cal...

Страница 21: ...s still unsuccessful check the camera exposure settings and adjust the exposure to optimize the quality of the images Use Best Exposure Setting procedure to let the system automatically determine the...

Страница 22: ...er of white and black squares are selected 10 Click on the Best Settings button iDimension 300 will find the best exposure setting for the environment The exposure setting value will update and the re...

Страница 23: ...sure set using best setting function This image is over exposed This image is under exposed The squares are slightly distorted in Image 1 as the paper was at a slight angle Once the paper is straighte...

Страница 24: ...d select Calibration Figure 3 16 Calibration Menu 2 Select the Calibration Settings Figure 3 17 Select the Calibration Settings Tab 3 Select the Zone of Interest tab 4 Press the Edit button to switch...

Страница 25: ...represents the detection area for an item to be dimensioned Figure 3 20 Work Area 7 Click on the orange Work Area button 8 Hold down the left mouse button and draw a rectangle around the work area Th...

Страница 26: ...Demo application to test the changes 1 Return to the QubeVu Manager main menu 2 Select Displays Figure 3 22 Select Demo from the QubeVu Manager Main Menu 3 Select Demo Displays 4 Place an item on the...

Страница 27: ...m x 5 cm Item position Dimensions only any position OCR Required any position with text and bar codes facing upwards Object colors All opaque packaging some variances may occur with glossy surfaces or...

Страница 28: ...er with Ethernet connection Customer applications can be integrated with iDimension 300 using a web service interface Configuration tool requires a JavaScript enabled browser Bar codes EAN 13 UPC A an...

Страница 29: ...PC being used Simply connect the unit to the PC using the Ethernet cable provided The QubeVu Manager tools will run via any compatible browser Open an Internet browser Enter the address or host name i...

Страница 30: ...the Restart button will restart the service that is running on the device Reboot will reboot the full operating system on the device Rebooting can take several minutes Running This device is running...

Страница 31: ...have been made while in the specific tool Some tools have multiple tabs selecting Cancel while viewing information on any tab will cancel edits made to all tabs Save The Save button will save all chan...

Страница 32: ...the Customer Display does not give the customer access to system level controls The Customer Display does not display images Figure 5 7 Customer Display System Information Button Zero Height Button Re...

Страница 33: ...changes It displays the dimensions of an item optional weight various status indicators and images Out of Bounds Indicator Indicates if the item has been placed within the viewable area Regular Shape...

Страница 34: ...ocess Log in with the username and password to access the QubeVu Manager Admin Tools Figure 5 8 iDimension 300 Manager Admin Tools Menu 5 3 1 Setup There are four parameters involved with the setup of...

Страница 35: ...32 iDimension 300 General Settings Figure 5 10 iDimension 300 General Settings Main Menu...

Страница 36: ...ge 21 the item is stable a stable weight has been received from a connected scale and if scale is present the scale weight has been settled Default Value On Valid Values On Off Flat Detection Determin...

Страница 37: ...scale Default Value 3000 Valid Values Scale Dependent Low Resolution Camera The Switch Resolution Delay ms is a minimum period of time in milliseconds that the system should pause after switching the...

Страница 38: ...ton to switch to the Configuring mode Valid Value Description grdsz Bar code map grid size integer in pixels dsfact Bar code map downsampling factor integer 1 psdn Psd score FFT size integer power of...

Страница 39: ...e and black squares Figure 5 11 Best Exposure Settings 3 Click on the Best Settings button Wait for QubeVu Manager to find the best exposure setting for the current environment The Exposure Setting va...

Страница 40: ...ution camera The longer the exposure time the brighter the image The value for exposure can be changed manually Use the Preview button to review results Default Value 100 Valid Values 1ms 255 ms Light...

Страница 41: ...e certification Figure 5 13 Date Time Tab Exposure setting using default value of 136 Image exposure set using best setting function In this case the default value was sufficient This image is over ex...

Страница 42: ...Figure 5 14 Change the Date 6 Click Done to save the settings 5 3 4 User Use the User screen to modify the password When selecting a new password keep these rules in mind Minimum length Six character...

Страница 43: ...terprise network Select Network from the setup menu There are two tabs Network Settings defines iDimension 300 as a network device in the enterprise network Network Security defines the security setti...

Страница 44: ...used to access the device from the Manager Tools in the future A hostname can be up to 15 characters For example http hostname IP Address If DHCP was checked an IP address will not be entered If DHCP...

Страница 45: ...ion object supplied with the device this setting will be either 76 or 77 mm depending on the version of the calibration object supplied Default Value 77 mm Valid Values 76 77 mm Target Finder Image Fi...

Страница 46: ...should account for vertical protru sions on the platform These are very rare cases when the platform may have a backstop or a lip Default Value Off Valid Values On Off Disk Finder Radius mm The radius...

Страница 47: ...alue to 1800 Depth Camera Height mm Distance in millimeters between the depth sensor and the base of the device Default Value 1530 Valid Values 600 1800 Parallax This is a read only field Correction f...

Страница 48: ...sion will be restarted after the restore Use the restore action from the Backup Menu to completely restore all settings To clear the parallax or to complete a factory calibration contact Rice Lake Cus...

Страница 49: ...ription Values Dimension Up A low pass filter is applied to the three dimensions of the object from every single frame that is captured This is intended to overcome noise and further improve accuracy...

Страница 50: ...Center Enabled Determines whether or not the system should relock based on the detection of a shift in the object s center Default Value Off Valid Values On Off Diff Center mm Specifies the amount of...

Страница 51: ...the cameras before it will be ready for use Use the restore action from the Backup Menu to completely restore all settings 5 4 4 Zone of Interest In general the Zone of Interest should not need to be...

Страница 52: ...e Calibration Settings tab Figure 5 23 Select the Calibration Settings Tab 3 Select the Zone of Interest tab 4 Press the Edit button to switch to configuration mode This will take a minute while the d...

Страница 53: ...changed A scale has been added or removed from operation To review the current camera calibration settings follow the steps below 1 Select Calibration from the Admin Tools menu 2 Select Camera Calibra...

Страница 54: ...move the calibration object until the center mark intersects with the yellow diagonal lines If the marks cannot be centered try to adjust the physical position of the head by adjusting the top sectio...

Страница 55: ...d the calibration object Figure 5 28 Draw a Rectangle Around the Calibration Object 11 Press Calibrate Figure 5 29 iDimension 300 Camera Calibration 12 All three tabs must show the green check mark fo...

Страница 56: ...autotrigger capture definitions for automatic triggering as required Changing the capture definition or defining a new capture definition is done by programmers when integrating with a client applicat...

Страница 57: ...n image High resolution image Bar code recognition There are four predefined profiles but it is possible to create new ones depending on the application processing requirements Figure 5 32 Capture Def...

Страница 58: ...ed Flat Do not dimension flats Parcel Do not dimension parcels All Neither flats nor parcels are dimensioned Low Res Camera Capture If checked it will create a low resolution camera image ResX and Res...

Страница 59: ...drive Figure 5 35 Select Firmware Version 4 Click Upload The firmware upgrade process will first copy the update file to the embedded processor It is recommended not to interrupt the upload process T...

Страница 60: ...ically restart to complete the upgrade While restarting the QubeVu Manager user interface cannot be displayed an error may display in the browser 8 Close the browser window 9 Restart the browser windo...

Страница 61: ...ckup Restore Screen 1 Select Backup from the Admin Tools main menu 2 Press to begin the backup process 3 Look to the bottom of the window for the file save dialog box To proceed with the backup select...

Страница 62: ...test hardware components and gather diagnostic information 5 6 1 Component Test 1 Click on Test All to test all hardware components Hardware components can be tested individually by clicking on the te...

Страница 63: ...tings of a serial scale attached to the system 1 Select the Serial Port and Scale Type from the drop down lists 2 Click the play icon to start the test Figure 5 41 Scale Test 3 If a valid scale is det...

Страница 64: ...sages The log view can be customized by Type all info debug or error Order Latest first or earliest first The log can be manually refreshed by clicking Refresh or automatically refreshed by ticking th...

Страница 65: ...the Information panel click on the Inspector button Device Information Figure 5 45 Device Information Tab The Device Information tab displays Model Serial number of this device Certification number t...

Страница 66: ...re 5 46 Change Counter The Change Counter displays the number of changes that have been done within the selected date range 1 To display all changes press the Search button without entering a date ran...

Страница 67: ...ure 5 48 Long Term Storage LTS is enabled from the Measurement Settings Screen It is disabled by default with a zero value To enable enter the appropriate number of days and click Save Viewing Long Te...

Страница 68: ...nt calibration file Figure 5 51 Measurement Settings 4 Click on the Delete Parallax button 5 Use the Restore function to restore from the factory settings 6 Re calibrate the cameras without the scale...

Страница 69: ...66 iDimension 300...

Страница 70: ...Administrator s Section 67...

Страница 71: ......

Страница 72: ...68 USA U S 800 472 6703 Canada Mexico 800 321 6703 International 715 234 9171 Europe 31 0 26 472 1319 www ricelake com www ricelake mx www ricelake eu www ricelake co in m ricelake com Rice Lake Weigh...

Отзывы: