Calibration
13
3.0
Calibration
3.1
Calibrate the Cameras
If using a scale, place the scale onto the base plate and put the calibration object on top of the scale. Center the
scale using the base plate marks. If not using a scale, place the calibration object directly onto the base plate. Center
the calibration object using the base plate marks.
Figure 3-1. Base Plate Marks
3.1.1
Define the Scale
If not using a scale, proceed to Section 3.1.2 on page 15.
1. Select
Setup
from the
Admin Tools
menu.
Figure 3-2. iDimension 300 Admin Tools Menu
2. Select
General Settings
from the
Setup
Menu.
3. Review the descriptions for scale type and scale comm parameters below.
4. Change the field values to match the scale.