![Rice Lake IDIMENSION 200 Скачать руководство пользователя страница 45](http://html1.mh-extra.com/html/rice-lake/idimension-200/idimension-200_installation-manual_1453986045.webp)
42
iDimension 300
5.4
Calibration
The following describes calibration information for
iDimension 300
.
5.4.1
Measurement Settings
Changes to the measurement settings will invalidate the Legal for Trade certification. Only change these
settings if Legal for Trade certification is not important to the enterprise.
Figure 5-17. Measurement Settings Main Menu
Calibration Parameters
Parameters
Description
Values
Platform Height
(mm)
This setting is primarily required when carrying out a factory calibration of the
cameras. The value specifies the height of any platform placed on top of the base
plate that is intended to be permanently located there (ie. a weigh scale). For a
factory calibration, this value must be set to ‘0.’
Default Value: 0 mm
Valid Values: <height of
platform>
Calibration Object
Height (mm)
The height of the object used to calibrate the device. If using the calibration object
supplied with the device, this setting will be either 76 or 77 mm depending on the
version of the calibration object supplied.
Default Value: 77 mm
Valid Values: 76-77 mm
Target Finder
Image File
Name of a bitmap file that describes the calibration pattern.
Default value:
RegistrationMark-
sCropped.bmp
Valid Values: Use default
file. Alternative files may be
supplied in the future.
Target Finder
Image Res (dpi)
Effective printed size of the calibration pattern on the calibration object. This size
is provided in terms of a resolution (in dots per inch) relative to the bitmap file that
was provided above.
Target Finder
Confidence (%)
As it searches for the calibration pattern, the system derives a confidence value
alongside its result. This parameter is the minimum confidence value required for
a calibration process to be deemed successful.
Default Value: 55%
Valid Values: 0-100%
Calibration Target
Finder Scale (%)
This value describes a tolerance for the expected size of the calibration pattern.
Default Value: 10%
Valid Values: 0-100%
Table 5-11. Calibration Parameters
Important
Содержание IDIMENSION 200
Страница 2: ...iDimension 300 Industrial Installation Manual PN 169040...
Страница 3: ......
Страница 15: ...12 iDimension 300 6 Click Done to save the settings...
Страница 35: ...32 iDimension 300 General Settings Figure 5 10 iDimension 300 General Settings Main Menu...
Страница 69: ...66 iDimension 300...
Страница 70: ...Administrator s Section 67...
Страница 71: ......