iDimension 300
Calibration not Successful
The calibration was not successful if any of the tabs show the orange warning sign. Try re-drawing the rectangle on
the
Depth
,
Low Resolution
and
High Resolution
tabs. Calibrate again until all three tabs have the green check mark
symbol.
Figure 3-12. Unsuccessful Calibration
If the calibration is still unsuccessful, check the camera exposure settings and adjust the exposure to optimize the
quality of the images. Use “Best Exposure Setting” procedure to let the system automatically determine the best
exposure settings for the environment.
1. Print the checkerboard pattern included in the Appendix on page 25.
2. Place the printed checkerboard pattern on the base (or scale).
3. Center the paper under the scanning head using the scale marks on the base as a guide.
4. Select
Setup
from the
Tools
menu.
Figure 3-13. Tools Menu
Image Quality Tab