482-AG Livestock Digital Weight Indicator
30
Visit our website
Streaming Paramaters for Local and Remote Operation
For systems that require two locations, local/remote support provides function equivalent to that of a Legal for Trade remote
display with keypad. Scale data from the local indicator is also displayed at the remote unit and keypad input from the remote
allows transactions to be initiated from either the local or remote unit.
To configure for local/remote operation, set
STREAM
to local or remote. Once the local unit scale has been configured, use the
SERIAL
menu or serial commands to set the serial parameters for local/remote operation. See
and
The printing operation can be performed at the remote unit and/or at the local unit. When the operator presses the
key at
the local unit, a printing ticket will be issued locally only. When the operator presses the
key at the remote unit, printing will
occur at the remote unit only.
BITS
8NONE
7EVEN
7ODD
The bits parameter determines the number of data bits and parity of data
SBITS
1 STOP
2 STOP
The stop bits parameter sets the number of stop bits to 1 or 2
TERMIN
CR/LF
CR
The termination character parameter sets the line termination character(s) for data sent
EOLDLY
000
0 – 255
The end-of-line delay parameter specifies, in 0.1 second intervals, the delay between transmitted lines of data
NOTE:
An end-of-line delay may be required for continuous transmission at slower baud rates to ensure the receiving buffer
is empty before another string is transmitted
ECHO
OFF
ON
The echo parameter specifies whether characters received by the port are echoed back to the sending unit
PRNMSG
OFF
ON
The print message parameter determines if
displays when a demand print is performed
STRUR
—
The stream update rate defines the data update rate when one of the stream formats is selected
STRLFT
When set to stream Legal for Trade, the weight information in the data stream is updated at the same rate as the indicator’s
display
STRIND
When set to stream industrial, the weight information in the data stream is updated at up to the A/D sample rate
Serial
Configuration
Parameter
Local Unit
Remote Unit
Parameter Value
Function
Parameter Value
Function
STREAM
Local
Stream Continuous Data
Remote
Sets indicator to display the weight
data sent from the Local Unit
COM2.TRIGER
Select STR-1
Streams the local unit weight
data to the remote unit
Select STR-1
Transmits demand key presses,
zero, units
COM2.BAUD
Local and remote must match
COM2.ECHO
OFF
OFF
Table 3-14. Local/Remote Configuration
Parameter Settings
Description
Table 3-13. COM2 Parameters (Continued)