Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
29
COM-1 Parameters
COM-2 Parameters
Parameter
Settings
Description
TRIGER
(
COM-1
)
—
The trigger parameter specifies the ports mechanism for triggering a transmission of data
DEMAND
When set to a demand port, printing is performed by pressing the
key or when a KPRINT EDP command is
received; this port also allows operation and EDP commands
PRN
When set to a printer port, printing is performed by pressing the
key; the port does not allow operation of
EDP commands
NONE
When set to
NONE
, the port is inactive
COMAND
When set to a command port, the port allows operation of EDP commands, but will not print
BAUD
9600
19200
38400
1200
2400
4800
The baud rate setting determines the transmission speed of data
BITS
8NONE
7EVEN
7ODD
The bits parameter determines the number of data bits and parity of data
SBITS
1 STOP
2 STOP
The stop bits parameter sets the number of stop bits to 1 or 2
TERMIN
CR/LF
CR
The termination character parameter sets the line termination character(s) for data sent
EOLDLY
000
0 – 255
The end-of-line delay parameter specifies, in 0.1 second intervals, the delay between transmitted lines of data
NOTE: An end-of-line delay may be required for continuous transmission at slower baud rates to ensure
the receiving buffer is empty before another string is transmitted.
ECHO
ON
OFF
The echo parameter specifies whether characters received by the port are echoed back to the sending unit
PRNMSG
OFF
ON
The print message parameter determines if
displays when a demand print is performed
Table 3-12. COM1 Parameters
Parameter Settings
Description
TRIGER
(COM-2)
—
The trigger parameter specifies the ports mechanism for triggering a transmission of data
NONE
When set to
NONE
, the port is inactive
PRN
When set to a printer port, printing is performed by pressing the
key; the port does not allow operation of EDP
commands
STR1
When set to
STR1
the port is used to transmit a continuous stream of data according to RLWS stream format**
STR2
When set to
STR2
the port is used to transmit a continuous stream of data according to Toledo8142 stream format**
STR3
When set to
STR3
the port is used to transmit a continuous stream of data according to Cardinal738 stream format**
STR4
When set to
STR4
the port is used to transmit a continuous stream of data according to Weightronix WI-120 stream format**
STR5
When set to
STR5
the port is used to transmit a continuous stream of data according to consolidated controls stream format**
**STREAM must be set to local to enable streaming
for more information on Stream formats.
Com-2 streams simultaneously on both the RS-232 and 20mA outputs, see
BAUD
9600
19200
38400
1200
2400
4800
The baud rate setting determines the transmission speed of data
Table 3-13. COM2 Parameters