6.1.2 Reporting Commands
Reporting EDP commands send specific information to the selected port. These commands can be used in both configuration
and weigh modes.
6.1.3 Special Function Commands
Reset Configuration
The
RESETCONFIGURATION
command can be used to restore all configuration parameters to their default values. Before
issuing this command, the indicator must be placed in configuration mode.
This command is equivalent to using the
DEFLT
function on the
TEST
menu, see
All load cell calibration settings are lost when the RESETCONFIGURATION command is run.
Soft Reset
The command
RS
will reset the indicator without losing configuration and calibration. It does reset consecutive number and
count mode.
6.1.4 Parameter Setting Commands
Parameter setting commands allow the operator to display or change the current value for a particular configuration parameter.
Current configuration parameter settings can be displayed in either configuration or weigh mode using the following syntax:
command<ENTER>
Parameter values can be changed only when the indicator is in the configuration mode using the following command syntax:
command=value<ENTER>
, where
value
is either a number or a parameter value.
Use no spaces before or after the equal (=) sign.
If an incorrect command is typed or the indicator is not in the configuration mode, the indicator responds with
??
.
Command
Function
DUMPALL
List all parameter values
DUMPAUDIT
List audit data information
P
Write current displayed weight with units identifier, see
ZZ
Write current weight and annunciator status, see
for more information.
S
Write one frame of stream format
VERSION
Returns the currently installed firmware version
Table 6-2. Reporting EDP Commands
Command
Description
Values
SC.GRADS
Graduations
1–100000
SC.ZTRKBND
Zero track band
0.0-3.0
SC.ZRANGE
Zero range
1–100
Example: to make it 5%, enter "SC.ZRANGE=5")
SC.INIZR
Initial zero range
0–100 (if set to zero, no initial zero is performed)
SC.MOTBAND
Motion band
0.0–100 (in display divisions)
Example: to enter 5 divisions, send
"SC.MOTBAND=5.0"
SC.OVRLOAD
Overload
FS+2%, FS+1D, FS+9D, FS
SC.DSPRAT
Display Rate
0.1 SEC,
0.25 SEC, 0.5 SEC, 0.75 SEC, 1 SEC, NONE
SC.SMPRAT
Sample rate
5HZ, 40HZ, 20HZ, 10HZ
SC.DIGFIL
Select Digital Filter Type
ADAPTIVE, ROLLING, OFF
SC.AFILTER.SENSITIVITY
Adaptive digital filter cutout sensitivity
LIGHT, MEDIUM, HEAVY
SC.AFILTER.THRESHOLD
Adaptive digital filter cutout threshold
0–999999
SC.RFILTER.SENSITIVITY
Digital filter cutout sensitivity
2OUT, 4OUT, 8OUT, 16OUT, 32OUT, 64OUT, 128OUT
Table 6-3. CONFIG EDP Commands