![Rice Lake 482-AG Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/rice-lake/482-ag/482-ag_technical-manual_1453829031.webp)
Configuration
© Rice Lake Weighing Systems
●
All Rights Reserved
27
3.6 Menu Parameters
Figure 3-8. Menu
3.6.1 Audit Trail Support
Audit trail support provides tracking information for configuration and calibration events. To prevent potential misuse, all
configuration and calibration changes are counted as change events.
Audit trail information can be printed by sending the
DUMPAUDIT
serial command. The audit trail display includes the legally
relevant (
LR
) version number (firmware version for the code that provides audit trail information), a calibration count and, if
REGULA
=
NTEP
or
CANADA
, a configuration count. The exact format of the information shown depends on the regulatory
agency specified for the
REGULA
parameter in the program (
PROGRM
) menu (
).
For
NTEP and Measurement Canada modes
, there are two numbers displayed: configuration (
CONFIG
) and calibration
(
CALIB
). The configuration number tracks changes to the legally relevant configuration parameter changes. The calibration
number tracks changes to the scale calibration.
In
OIML Mode
, there is only one number displayed; Audit Trail. The Audit trail number is incremental for a legally relevant
configuration parameter changes and all calibration changes. The last date the scale was calibrated is also displayed
Parameter
Settings
Description
AUDIT
CALIB
CONFIG
View audit trail including the time and date of last configuration or calibration; options: CNT, TIME, DATE
UNITID
0–999999
View and edit the Unit ID (up to a six digit number)
TIMDAT
TIME
DATE
View/set time and date as currently programmed in the indicator; format date as
MMDDYY, DDMMYY, YYMMDD
or YYDDMM
SERIAL
COM-1
COM-2
STREAM
Configure serial ports, see
PFRMAT
GFMT
GWFMT
LTFMT
View or configure the print format used for gross, group weight and load total print tickets;
see
VERS
FIRMW
LR
Displays the firmware and legally relevant firmware currently installed
MISC
POWER
BKLGHT
BAT
Set miscellaneous parameters; see
Table 3-10. Menu Parameters
SERIAL
PFRMAT
MISC
AUDIT
UNITID
MENU
CONFIG
FORMAT
CALIBR
PROGRM
PASWRD
TEST
VERS
TIMDAT
See
See
See
See
Содержание 482-AG