5.9 Purging and disinfecting the WD
All parts of the machine supplying fluids to the load or chamber are purged and disinfected during
a normal operating cycle. The wash chamber and pipework of the machine have been designed
so that any remaining process water shall flow towards the discharge point of the machine.
Should the machine not have been used for a period longer than 24 hours, it is recommended
that a short cycle including disinfection be run before placing a load into the machine for
processing. However, all the cycles validated and sealed by the manufacturer include a
disinfection phase as the last phase of the process.
Service engineers shall pay attention to
the state of the machine before dismantling any parts of the chamber or pipework. Unless
informed otherwise, it shall always be assumed that there may be harmful
microorganisms present by a Fault condition or water system breakdown.
5.10 Methods of cleaning
Daily:
Check that all the sprayers rotate freely and the spray holes are open
In case the fixed or rotating nozzles get blocked the cleaning effect will deteriorate. Please take
care of keeping the nozzles clear
Weekly:
Use a toilet brush and clean the outlet sieve by brushing it a few time
Clean the outer stainless steel panels with diluted washing agent and soft washcloth; flush,
wipe and dry. See also chapter 10.3 (Maintenance manual)
Chamber cleaning: see section 8.6 (Maintenance manual)