5.7.6 Malfunctions
WARNING situations, which can be controlled by the User, are being indicated as shown below.
If the recommended actions have been taken, press the
“CONTINUE”
key .
After the fifth indication (door open or lack of chemical) the machine will switch over into FAULT
condition.
See warnings from the table below:
Warning text
Explanation
Chemical tank 1 empty
Change or re-fill the detergent container 1
Chemical tank 2 empty (Option)
Change or re-fill the detergent container 2
Loading door is open
The loading door is open or not closed
correctly
Chamber too hot for starting the cycle.
Wait or open the door for cooling the
chamber.
Or press cancel to stop the cycle"
A new cycle will not start if the chamber
temperature is over 60
0
C
No disinfection
Must be confirmed by the user if the cycle
does not include disinfection phase.
Cycle No. # OK, No disinfection
Must be confirmed by the user if the cycle
does not include disinfection phase.
Open water valves
Water valves shall be open when starting.
Power cut during the cycle
There has been a power cut and
the ongoing cycle must be restarted
Periodic service is needed
Will appear if this feature has been selected
Set parameters
Contact service / Deko distributor
Check machine date and time
Contact service / Deko distributor
USB memory card is missing
Insert the USB card
Error while saving report
Check SD-memory card"
Contact service / Deko distributor
FTP- file transfer failed
Contact service / Deko distributor
Communication error
Contact service / Deko distributor
Error while printing report
Contact service / Deko distributor
Check calibration values
Contact service / Deko distributor
P
Power cut off during the cycle
Re-Start the cycle