Message
Cause(s)
Solution
LQC Lockout. Ejecting.
LQC is overdue for this Assay and no
overrides are available.
LQC must be run and passed, or a full
access user can reset the LQC timer
and/or override for this Assay in
Reader Settings on the RCU.
Invalid Assay or Control, Aborting LQC.
An incorrect Assay has been selected to
run LQC.
Repeat LQC with the correct Assay or
contact the Supervisor.
LQC required. Assay Aborted. Ejecting.
LQC has expired and all available
overrides have been used. Tests cannot
be run until the required LQC has been
run and passed.
Run the required LQC.
Message
Cause(s)
Solution
! All Databases Will Be Erased
Back up files at Settings: Export
Configuration
System will restart following this
procedure
Request was made to upgrade software
or reset system to default settings.
Cancel if not intended.
Before performing this operation, all
databases should be backed up using
Export Configuration.
! All Results Will Be Deleted
Request was made to delete all results.
ERROR - Missing Firmware Version File
The file required to complete the
upgrade is missing.
Contact Technical Support
ERROR - Upgrade Failed
Restart to Try again
Upgrade failed.
Restart the RAMP
®
200 and re-try
upgrade. If the problem persists,
contact Technical Support.
Содержание RAMP 200
Страница 7: ...A ...
Страница 13: ...B ...
Страница 15: ...C ...
Страница 20: ...Figure 4 RAMP 200 Test Module rear panel label ...
Страница 21: ...D ...
Страница 25: ...E ...
Страница 31: ...F ...
Страница 37: ...G ...
Страница 44: ...H ...
Страница 54: ......
Страница 55: ...I ...
Страница 62: ...J ...
Страница 65: ...K ...
Страница 73: ...L ...
Страница 85: ...M ...
Страница 94: ......