Do not use excessive force when inserting the Test Cartridge. Once inserted, the TM will
automatically guide the cartridge into position for analysis.
NOTE
The Test Cartridge must be inserted into the TM within
approximately 30 seconds of sample application; otherwise
FAIL
will be displayed in the port icon and the system will abort the
Assay.
If the Test Cartridge is not inserted within five minutes of the
Add Sample & Insert
display
prompt, the instrument will time out. Touch
OK
to return to the
Main Menu
.
6
Once the
Cartridge Inserted
message clears, the display returns to the previous screen and
the active port icon
shows a timer that counts up until sample flow is detected by the
RAMP® 200. The timer then counts down until the RAMP® Test is complete.
7
While the Assay is in progress, touching the port icon will show details of the Assay
progression status and an
Abort Assay
option.
Selecting
Abort Assay
will eject the Test Cartridge and the RAMP® 200 will record the Test
as a Failed Assay in the system
Events Log
.
To return to the previous screen, touch
Exit
.
NOTE
Selecting
Run Assay
or
Run LQC
will start another Assay/LQC and the
Test sequence will begin again using a different available port. The
Assay progress can be checked by touching the port icon. Touch
Exit
to
return to the previous screen.
8
When the Assay is complete, the Test Cartridge is automatically ejected and the
corresponding port icon will display
.
To view the Test details, touch the icon for that port prior to physically removing the Test
Cartridge.
9
Remove the Test Cartridge from the TM test port.
Содержание RAMP 200
Страница 7: ...A ...
Страница 13: ...B ...
Страница 15: ...C ...
Страница 20: ...Figure 4 RAMP 200 Test Module rear panel label ...
Страница 21: ...D ...
Страница 25: ...E ...
Страница 31: ...F ...
Страница 37: ...G ...
Страница 44: ...H ...
Страница 54: ......
Страница 55: ...I ...
Страница 62: ...J ...
Страница 65: ...K ...
Страница 73: ...L ...
Страница 85: ...M ...
Страница 94: ......