6
•
The
S8 C
OMPACT
can be set to deliver pressures up to 20 cm H
2
O. In the unlikely
event of certain fault conditions, pressures up to 30 cm H
2
O are possible.
•
If oxygen is used with this device, the oxygen flow must be turned off when the
device is not operating.
If oxygen has been left on, turn off the flow generator, then wait 30 minutes
before turning on the flow generator again.
Explanation:
When the CPAP device is not in operation and the oxygen flow is
left on, oxygen delivered into the air delivery tubing may accumulate within the
CPAP machine enclosure and create a risk of fire. This applies to most types of
CPAP machines.
•
Oxygen supports combustion. Oxygen should not be used while you are
smoking or in the presence of an open flame.
•
Always ensure airflow is being generated by the device before the oxygen
supply is turned on.
•
Always turn the oxygen supply off before stopping the airflow from the device.
Note:
At a fixed rate of supplemental oxygen flow, the inhaled oxygen concentration
will vary, depending on where the oxygen is introduced, the pressure settings, patient
breathing pattern, mask selection, and leak rate.
•
Do not use the
S8 C
OMPACT
if there are obvious external defects or unexplained
changes in performance.
•
Do not open the
S8 C
OMPACT
case. There are no user serviceable parts inside.
Repairs and internal servicing should only be performed by an authorized service
agent.
Note:
The plate on the back of the device is a non removable cover. Removal of this cover
voids warranty.
•
Explosion hazard—do not use in the vicinity of flammable anesthetics.
C
AUTIONS
•
At low pressures, the flow through the exhalation ports of your mask may not
clear all exhaled gas from the tubing. Some rebreathing may occur.
•
The air flow for breathing produced by this device can be as much as 11°F (6°C)
higher than the temperature of the room. Caution should be exercised if the room
temperature is warmer than 90°F (32°C).
Note:
The above are general warnings and cautions. Specific warnings, cautions, and
notes appear with the relevant instructions in the manual.
338251r1.book Page 6 Friday, June 10, 2005 5:03 PM
Содержание S8 Compact System
Страница 9: ...4 338251r1 book Page 4 Friday June 10 2005 5 03 PM...
Страница 13: ...8 338251r1 book Page 8 Friday June 10 2005 5 03 PM...
Страница 31: ...26 338251r1 book Page 26 Friday June 10 2005 5 03 PM...
Страница 40: ...35 MANUEL CLINIQUE Fran ais S8 Compact F RAN AIS 338251r1 book Page 35 Friday June 10 2005 5 03 PM...
Страница 43: ...38 338251r1 book Page 38 Friday June 10 2005 5 03 PM...
Страница 47: ...42 338251r1 book Page 42 Friday June 10 2005 5 03 PM...
Страница 65: ...60 338251r1 book Page 60 Friday June 10 2005 5 03 PM...
Страница 74: ...69 MANUAL CL NICO Espa ol Sistema S8 Compact E SPA OL 338251r1 book Page 69 Friday June 10 2005 5 03 PM...
Страница 77: ...72 338251r1 book Page 72 Friday June 10 2005 5 03 PM...
Страница 81: ...76 338251r1 book Page 76 Friday June 10 2005 5 03 PM...
Страница 99: ...94 338251r1 book Page 94 Friday June 10 2005 5 03 PM...
Страница 108: ...103 MANUAL CL NICO Portugu s Sistema S8 Compact P ORTUGU S 338251r1 book Page 103 Friday June 10 2005 5 03 PM...
Страница 111: ...106 338251r1 book Page 106 Friday June 10 2005 5 03 PM...
Страница 115: ...110 338251r1 book Page 110 Friday June 10 2005 5 03 PM...
Страница 133: ...128 338251r1 book Page 128 Friday June 10 2005 5 03 PM...