English 29
Data storage
The AirSense 10 device stores patient compliance data such as AHI, Total Hours Used and Leak.
Detailed Data such as snore and pulse rate, as well as high resolution flow and pressure data, are
stored on the SD card.
Data can be transmitted to therapy management software either remotely via cellular
communication, or via SD card. The different ways of transmitting data are detailed in the table
below.
For more information on compliance management with AirView or ResScan, refer to the manuals
supplied with the software.
Type of data
Transmission method
Sessions stored
Cellular
communication
to AirView
SD card to
ResScan
SD Card to
AirView
(card-to-
cloud)
Summary data (compliance data)
365
Detailed data
Limited by usage and
SD card storage
capacity
High resolution flow and pressure data
(25 Hz)
Detailed data are stored on the SD card and can be viewed on ResScan. Examples of detailed data
available are shown below.
Detailed Data
Parameter Sampling
rate
Apnea or hypopnea events
Aperiodic
CSR Aperiodic
RERA (AirSense 10 AutoSet for Her only)
Aperiodic
Flow limitation (flat to round)
1/2 Hz
Leak (L/sec)
1/2 Hz
Minute ventilation (L/min)
1/2 Hz
Pressure (cm H
2
O / hPa)
1/2 Hz
Snore (quiet to loud)
1/2 Hz
Pulse rate (beats/min)—if an oximeter adapter is attached
1 Hz
Oxygen saturation (SpO
2
)—if an oximeter adapter is attached
1 Hz
Software upgrade
The device has a software upgrade feature. When a software upgrade is in progress, the screen will
flash for approximately 10 minutes.
Содержание AirSense S10
Страница 1: ...Clinical guide English ...