XCell™ C410:V3 Controller
User Guide
36 repligen.com UG-C410v3-06
Figure 20. Algorithm screen
XCell ATF® Device cycle change can occur by either of three algorithms. The screen tracks cycle
change Algorithm performance. The three Algorithms are:
1.
Set Point - normal process based on set up parameters.
2.
Overtime - when P2 Set Point is not achieved in within 120% of calculated time.
3.
Overflow - When P2 reaches P1 (+/- OSP) within the set delay time.
Table 17. Batch Info Algorithm display parameters
Field
Description
Algorithm performance
Tracks cycle change mechanism
Primary method
Shows cycle count triggered by Set point method, tracks pressure cycle and
exhaust cycle counts
Overtime
Shows cycle count triggered by overtime, tracks pressure cycle and Exhaust
cycle counts
Overflow
Shows cycle count triggered by overflow, tracks pressure cycle and Exhaust
cycle counts
Button
Description
Reset count
Resets all counts, by Set Point, Overtime and Overflow to zero
Overview
Navigates to Batch Overview screens
All Primary
All Primary screen buttons are displayed to navigate to those screens