1-7
BATTERY POWER
DO NOT ALLOW COOLANT
OR DEBRIS TO ENTER THE
BATTERY COMPARTMENT
Tighten
The O-M-I or MI 12 interface low battery LED will
light up when the battery needs to be replaced.
This must be done as soon as is practicable
following a gauging cycle.
To replace the battery, use a 3 mm hexagon key
to remove the battery cap. If battery cap is not
accessible, remove the OMP from the machine.
When replacing the battery insert the battery
+
end
first, then replace the cap. The battery cap screw
torque value is 1 Nm (0.74 lbf.ft).
If the OMP has been removed from the machine,
replace the OMP and recalibrate the probe system,
before commencing the next gauging cycle.
65 days
35 days
STAND-BY
LIFE
5% USAGE
72 min/day
TYPICAL BATTERY LIFE
Duracell lithium battery type DL123A
CONTINUOUS
LIFE
88 hours
Optical on
Timer off
Optical on
Timer off
Battery booster module
3V lithium battery
End cover
3mm hexagon key