1-12
FAULT FINDING - If in doubt, consult your probe supplier.
POOR PROBE REPEATABILITY
Debris on part.
Clean part.
Turret
Verify probe repeatability
repeatability poor.
using single point move.
Loose mounting of
Check and tighten as
probe/loose stylus.
appropriate.
Calibration and update
Review probe software.
of offsets not occuring.
Calibration and probing
Review probe software.
speeds not the same.
Calibrated feature has
Check position.
moved.
Measurement occurs as
Reveiw probe software.
stylus leaves surface.
PROBE CRASHES
Inspection probe using
When two systems active,
tool setting probe
isolate tool setting probe.
signals.
Workpiece obstructing
Review probe software.
probe path.
Probe length offset
Review probe software.
missing.
PROBE STOPS IN MID-CYCLE
Beam obstructed.
Check O-M-I/MI 12
error LED.
Remove obstruction.
Probe collision.
Find cause and rectify.
Damaged cable.
Check cables.
Power supply lost.
Check power supply.
Probe unable to find
Part missing or out of
target surface.
position.
PROBE FAILS TO SWITCH ON
Probe is already
If necessary switch
switched on.
probe off.
Dead battery.
Change battery.
Battery installed
Check battery installation.
incorrectly.
Probe not properly
Check alignment and if
aligned with OMM/O-M-I. OMM/O-M-I fixing is secure.
Beam obstructed.
Check if OMM/O-M-I window
is clean/remove obstruction.
OMM/O-M-I signal
See performance envelope.
too weak.
See pages 1-4 and 1-5.
No O-M-I start signal
See page 1-18.
No power to MI 12
Check if stable 24 V supply is
or O-M-I
available.
Check connections and fuses.