R0E521000CPE00 User’s Manual
2. Setup
REJ10J0845-0600 Rev.6.00 Aug. 20, 2009
Page 36 of 92
2.8.2 If an Error is Detected in the Self-check
Table 2.7 lists how to remedy the troubles if the target status LED display is abnormal in the self-check. When an error is
detected, shut off the emulator and the user system and follow the steps in the Table 2.7.
Table 2.7 Error display in the self-check and how to remedy it
LED display
Blinking
OFF
POWER CLOCK RESET RUN
Problem & Remedy
The emulator system is not working properly.
- Check that power is supplied to the emulator.
- The emulator may be damaged. Contact your local distributor.
A clock is not supplied to the emulator.
- Check that the oscillator circuit board (OSC-3) is attached.
The power is not supplied to the emulator.
- Check that the power supply cable is connected properly.
- Check of jumper settings (see Table 2.6).
The emulator system is not working properly.
- The emulator may be damaged. Contact your local distributor.
IMPORTANT
Notes on the Self-check:
z
Be sure to disconnect the user system
before executing the self-check.
Use the preinstalled oscillator circuit board OSC-3 (20 MHz) to execute the self-check.
z
If the self-check does not result normally (excluding target status errors), the emulator may be damaged. Then,
contact your local distributor.