49
15.4.
E
LECTRICAL DATA
MCL II control box
Voltage in:
100 – 240V, 50/60 Hz
Current in:
Max. 3,15A
Electric shock protection:
Class II
Duty cycle:*
10%
2 mins
of continuous use followed by
18 mins
not in use
5 switching cycles per minute
*Electrically operated beds are intended to be operated intermittently rather than continuously. If the
bed is operated continuously for up to 2 minutes it must then be left for at least 18 minutes before reuse
to allow the electrical system to cool sufficiently. If the bed is continuously used for an extended period
of time and it exceeds the duty cycle the control box may become temporarily disabled or irreparably
damaged.
No more than two drives may be operated at rated load simultaneously!
Safety standards:
IEC 60601-1: 2005
IEC 60601-2-52:2009
IEC 60601-1-11:2010
Applied part electrical shock
protection:
Type BF
Applied parts:
Mattress platform
Profiling sections
Bed ends
Handset
Side rails
Liquid ingress protection:
IPX4 – Splash resistant
Noise level:
67dB(A)
Service life:
5 years
Environmental conditions:
Operational Limits*
Transportation/storage limits
Ambient temperature:
+10ºC to +40°C
-20°C to +50ºC
Humidity:
30% - 75%
30% - 75%
Atmospheric pressure:
800 to 1060 hPa
800 to 1060 hPa
Altitude:
≤ 2000 m
≤ 2000 m
* Always ensure the bed is brought up/down to room temperature before plugging in and operating. It
is recommended to leave the bed for at least 2 hours in order to ensure that it reaches room
temperature.