Menu
(use keys [2] or [8])
Setting value
(use keys [4] or [6])
(*=default)
Description
Auto
on*/oFF
Auto dispense
MAtCh
on/oFF*
Match mode
bLE
see
Bluetooth Setting
for more details
nAME
Composed of 5 digits (0~9)
AB
Set bluetooth parameters AB
(
default AB = 00)
PoWdE
see pg.14,
User Defined Powder Configuration
for more details
uSr
Configure user defined powder
configuration 06~20
A
Powder Size 1~4
B
Standard/match mode: 0=match,
1=standard
C
Weight unit: 0=gm, 1=gn
D
Minimum operation value for large
tube: 1~40
E
Expected large tube stop point:
1~99
F
Expected small tube stop point:
0~99
G
G
1
G
2
Angle and response time for
short trickle: 0,0~9,9
H
Expected start point for short trickle:
0~99
I
I
1
I
2
Angle and
response time for
short trickle: 0,0~9,9
Sound
on*/oFF
Buzzer sound
7
CALIBRATION
Always warm up the scale 15-20 minutes before calibration. Always calibrate your scale before
use. This must be done
WITHOUT
the powder pan on the platen. To calibrate the scale, follow
the steps below:
1. Be sure the powder pan has been removed from the scale platen. When you turn the
scale on, the display will show 99999 across the display for a few seconds and then counts
down to display 0.0 GN (grains mode). Once “Stable” is displayed, the scale is ready to be
calibrated. See
Fig. 1 (next page)
. Note: The unit will automatically switch to STD (standard)
mode (Match mode "oFF") when the calibration process is initiated.
2. Next please press [CAL] key on the scale. The display will show C 0, and "GM" will be
displayed in the lower right hand corner of the display. See
Fig. 2
.
3. When the display shows C 0 and “Stable” shows up, press the [CAL] key again.
4. When the display shows C 20, place a 20 gram weight on the platen and when “Stable” is
displayed, press the [CAL] key. See
Fig. 3
.