7
4. Calibration Device LSG and LSK
•
Before calibration fill up the wash bottle
with water (up to middle position
between „MIN.“ and „MAX.“)
•
the very left connection is for the
humidity probe and both right
connections are for the instruments
which should be calibrated
humidity probe
•
connect gas bottle (LSG with 1000ppm Methane in synth. air, LSK with R134A in
synth.air)
•
adjust the left flow meter up to one (gas flow)
•
the humidity probe is adjusted with the right flow meter (45- 55 % rF synthetic air)
•
then follow the calibration procedures