2.3  Scale of the bar graph display  

 

  1 Bar 

:          0 ppm  

 

  2 Bars  

: >     40 ppm  

 

  3 Bars  

: >     80 ppm  

 

  4 Bars  

: >   125 ppm  

 

  5 Bars  

: >   175 ppm  

 

  6 Bars  

: >   250 ppm  

 

  7 Bars  

: >   350 ppm  

 

  8 Bars  

: >   500 ppm  

 

  9 Bars  

: >   700 ppm  

 

10 Bars  

: >   900 ppm  

 

11 Bars  

: > 1150 ppm  

 

12 Bars  

: > 1500 ppm  

 

13 Bars  

: > 3000 ppm  

 

14 Bars  

: > 4500 ppm  

   
These  data  are  just  approximate  values  and  are  not  for  end  users  information.  Since  the 
LSG/LSK are calibrated with calibration gas at 1000 ppm following statement can be made:  
at 1000 ppm the display will be between 9 and 11 bars!   
 

2.4 Sensitivity   

The sensitivity can be set to “1/5“ (left bars blinking), “normal” (middle bars blinking) and  
“factor 5” (right bars are blinking)..   
At „factor 5“ ten bars will be shown already at a concentration of 200 ppm.  
    

2.5 Automatic Switch Off  

Should no button be pressed the instrument will switch off after 20 minutes.  
 

2.6 Battery  

At all display mode the grey battery symbol will be shown once the battery voltage is be below 
2.10V. Below 2.0V the battery symbol will be shown as black.  
When the battery voltage is below 1.9V the instrument will switch off..  
The operating time without lightning is at approximately 16 hours and at approximately 12 hours 
with lightning. 
 
It  appears  that  the  contact  resistance  between  the  batteries  and  the  clip  contacts  have  a 
negative impact on the operating time. . If might be useful to turn the inserted batteries in the 
clip contacts several times. 
 

Содержание ecom LSG

Страница 1: ...von Carsten Pauli E Mail pauli rbr de Erstellungsdatum 22 02 2011 rbr Messtechnik GmbH Am Gro en Teich 2 D 58640 Iserlohn Tel 49 0 2371 945 5 Fax 49 0 2371 40305 Internet www rbr de E Mail info rbr de...

Страница 2: ...phase 4 2 2 Measurement 4 2 3 Scale of the bar graph display 5 2 4 Sensitivity 5 2 5 Automatic Switch Off 5 2 6 Battery 5 2 7 Operation via the Jack Plug 6 3 Calibration 6 3 1 Change in the software f...

Страница 3: ...O button the instrument is switched on and the display mode can be changed By long pressing of the I O button the LSG LSK will be switched off Press the button as long as the bars cannot be seen anym...

Страница 4: ...the consumption increases to 180mW 20mW Attention The energy consumption is pulse shaped 1 7 Beep The beep will sound for function control during switch on The ppm beep can be switched off optionally...

Страница 5: ...when the instrument is switched off It is done when the instrument is switched on a warm up phase will be run automatically During the first 10 seconds of the warm up phase the voltage consumption of...

Страница 6: ...ng and factor 5 right bars are blinking At factor 5 ten bars will be shown already at a concentration of 200 ppm 2 5 Automatic Switch Off Should no button be pressed the instrument will switch off aft...

Страница 7: ...zero calibration The detectable gas may not be in the ambient air The zero calibration will be done automatically after every warm up phase The resistance of the sensor must be in the range of 0 333 t...

Страница 8: ...the humidity probe and both right connections are for the instruments which should be calibrated humidity probe connect gas bottle LSG with 1000ppm Methane in synth air LSK with R134A in synth air ad...

Страница 9: ...son Failure reason Failure reason Failure reason Bar 14 wrong calibration when making the zero calibration a gas concentration was present to less difference between zero and 1000ppm calibration Senso...

Страница 10: ...is switched Instrument Instrument Instrument Instrument Sensor Sensor Sensor Sensor app sensor resistance app sensor resistance app sensor resistance app sensor resistance in air in air in air in air...

Страница 11: ...7 Testing Bottle A mixture is made of 40 parts water and one part Ethanol Ethanol 652 EG No 200 578 8 Of this mixture 8 ml will be filled in the cotton of the bottle The closing of the bottle has to...

Страница 12: ...tly to the Cyclone Pro and contains all settings When programming t When programming t When programming t When programming the sensor head may not be populated he sensor head may not be populated he s...

Страница 13: ...button menu than press the start button on the Cyclone Pro transfer starts 9 Exchange the Sensor unit The customer and all service department could exchange the sensor unit The sensor unit is pre cali...

Страница 14: ...ion Even after the warm up phase is finished a new zero point is determined cyclical and the linearization is calculated new when having a positive sensor drift resistor bigger than to the end of the...

Отзывы: