84
RBCBioscience
1. Enter the vender page, and key in the
password.
2. Enter optic parameter page.
• If the count value sre
higher
than the range,
decrease
the integration time (1 ms).
• If the count value sre
lower
than the range,
increase
the integration time (1 ms).
• Press "Integration Time" to adjust the time,
and press"Finish to complete the adjustment.
3. Press “Update” to finish the adjustment.
4. Note that Empty cuvette test should be
performed after change integration time.
Adjust integration time
Wavelengh
Tolerance
260mm
±750
280mm
±1250
320mm
±1250