75
1. Select Maintenance.
2. Select Vender and input password.
3. Select Teach Position and enter the teaching
page.
4. The wiring configuration of the connector of
the motor on the main board.
Disconnect X ,Y , C Motor Power Cable
The pre-operation before
calibration
Start
Maintenance
Barcode
Optics Data
Prev
Maintenance
Update Program
Vender
Piercing Needles
System Status
Optic Test
Alarm History
Back
X
V
Y
C
Содержание MagCore HF16 Plus
Страница 1: ...MagCore Super HF16Plus EngineerInTraining manual ...
Страница 34: ...33 Disassemble thebackservicelid Disassemble back service lid ...
Страница 35: ...34 RBCBioscience Disassemblethe topservicelid Po ...
Страница 45: ...44 RBCBioscience ReplaceDoorSensor 1 Open the front door 2 Pull out the sensor 3 Disconnect the connector ...
Страница 63: ...62 RBCBioscience Maintenance Electricalcontrolandengineeringsoftware ...
Страница 66: ...65 Mainboardconfiguration ...
Страница 75: ...74 RBCBioscience PositionTeachingTools 200µlSPTip toolA toolB toolC PipetteTip ...
Страница 87: ...86 RBCBioscience ...
Страница 93: ...92 RBCBioscience ...