7
━ IQ Documents
IQ/OQ/PQ Documents
MagCore
Super and HF16 Plus
Installation Qualification/Operation Qualification/Performance Qualification
Serial No.____________________Date____________
Installation Qualification:
IQ check list
PASS
FAIL
NOTE
Machine Inspection:
* No obvious defect on casing.
* Check handle.
* No obvious defect on PMMA LOGO surface.
* Check T‐rack.
* Check Cartridge rack.
* Check door function.
□
□
□
□
□
□
□
□
□
□
□
□
Function Test:
* Check outlet voltage:110~220V.
* Turn on machine.
* Check 7 inch touch panel.
* Check LED LOGO light.
* Barcode scan function
(Not for HF16 Plus)
.
* UV lamp Test.
* Motor Test (All Axis Test).
* Reject Tip Test (Act smoothly).
* Heater Test (RT to 60
C less than 7mins).
* System Time Setup (In System Status).
□
□
□
□
□
□
□
□
□
□
□
□
□
□
□
□
□
□
□
□
□
□
Examination
Signature
Verification
Signature
Page 1 of 3