━ Install Tube, Tip
Warning :
Please do not use the Tips and Tubes which are not provided by the original manufacturer.
The test result may be not correct and the machine may be damaged due to different Tips
and Tubes.
━ Install reagent cartridge
━ Install MagCore® Cuvette
(MagCore Super Only / for optic test)
1. Please put MagCore® Cuvette into the
corresponding sample's well. The installation
of Cuvette is directional. It is impossible to put
it into the well if the direction is not correct.
2. Put the 200 μl SP tip into the W4 of the T-Rack.
Please insert the front end of the cartridge into the space
below the fixing plate of the Cartridge Rack.
1. Put the tip into the corresponding well according to the left
figure.
2. Put the T-Rack on the machine.
Please install the Tip and Tube
according to the instructions of
extraction kit user manual.
Please insert the Cartridge Rack before
the T-Rack.
W5 Elution Tube
W4 200μl SP Tip
W3 Pipette Tip
W2 DNase / Tip Holder set
W1 Sample Tube - FS Tip
When installing it,
please do not scratch
the optic test window.
Содержание MagCore HF16 Plus
Страница 1: ...MagCore Super HF16Plus EngineerInTraining manual ...
Страница 34: ...33 Disassemble thebackservicelid Disassemble back service lid ...
Страница 35: ...34 RBCBioscience Disassemblethe topservicelid Po ...
Страница 45: ...44 RBCBioscience ReplaceDoorSensor 1 Open the front door 2 Pull out the sensor 3 Disconnect the connector ...
Страница 63: ...62 RBCBioscience Maintenance Electricalcontrolandengineeringsoftware ...
Страница 66: ...65 Mainboardconfiguration ...
Страница 75: ...74 RBCBioscience PositionTeachingTools 200µlSPTip toolA toolB toolC PipetteTip ...
Страница 87: ...86 RBCBioscience ...
Страница 93: ...92 RBCBioscience ...
Страница 94: ......
Страница 95: ......