OPERATION:

8

0

AUTO (Automatic)

The AUTO (Automatic) mode will measure and store, automatically, numerous single sequential
tests at timed intervals that are preset by the operator. To preset the number of tests at timed
intervals please go to the Set-UP screen (See below) and select AUTO.
In this AUTO mode the instrument will continuously measure multiple tests until the operator
selects and presses STOP or until the number of tests that were preselected have been achieved.
The instrument will automatically save test results in a numbered memory location that will
increment by a timed interval.

AUTO Setup screen:
Ti = Test Interval
= Number of Tests desired.

A maximum of 10 minutes at 5 second intervals can be preset.

Measurement Notes:
Note 1: The instrument will halt the measuring process if excessive interference is detected.

Note 2: If a warning message,”Rx Too High” is displayed, Or if ”+++.+” is displayed, Then
return to the “SET UP” screen and select the next higher current range.
The ”+++.+” is displayed when no input is measured on the +P &

- P terminals.

ARCHIVE:

9

0

The Archive or Memory allows over 2000 complete test results to be saved. The method used for
the memory location is:

Each memory location is designated a number, a date and time on which it was recorded, the
results measured, and the current it was tested at.

If multiple tests are taken in one test sequence the memory location will appear like this:

The test results can be printed from the Archive via the RS232 port or Parallel printer port.

Содержание MC2

Страница 1: ...RAYTECH USA Inc 118 S 2nd Street Perkasie PA 18944 USA Tel 888 4 THE SPY Tel 267 404 2676 Fax 267 404 2685 200 Amp Digital Micro Ohm Meter Model MC2 Instruction Manual Ver 2 0...

Страница 2: ...tails 5 0 7 Limitations of Testing 7 Front Panel Description 6 0 8 Typical Display screens 7 0 10 Main Selection Archive Test 10 Store Print Auto Test Clock Contrast 11 Getting Started 8 0 12 Operatio...

Страница 3: ...em The MC2 is line operated The system operates from an auto sense input power from the main power input from 100 to 250 vac 50 60Hz DO NOT OPERATE IN AN EXPLOSIVE ATMOSPHERE Do not operate the instru...

Страница 4: ...Ground of paper If any of the above items are missing or damaged contact your local representative or Raytech USA Inc immediately The test leads may be ordered other than shown RAYTECH Toll free serv...

Страница 5: ...p 0 0 5 0 0 1 Rdg 0 2 5 Digits or 0 2 TEMPERATURE Operating 10 C to 60 C Storage 20 C to 70 C CABLE SET ACCESSORIES Current and potential lead set 5 meters Power cord Heavy duty 7 6 meters Instruction...

Страница 6: ...e of the most lightweight systems available that comes complete with its own rugged waterproof Fieldcase Simple Maintenance There is No maintenance required There is No calibration procedure No potent...

Страница 7: ...g require a long period to saturate the winding to get a stable reading In addition care must be taken when removing test leads such that a voltage charge does not remain on the winding Lethal current...

Страница 8: ...250 Vac Press to stop Turn Clockwise to release Computer Connections Printer Serial Port RS232 Bi directional The panel mount printer uses Female 9 pin connector See Section 10 thermal heat transfer t...

Страница 9: ...perature Probe connection 2 pin Connection for optional external temperature probe Buttons This is a multi actuation rotary knob which is the main user input and control This rotary knob allows the us...

Страница 10: ...MAIN SELECTION screen START TEST TEST IN PROGRESS END OF TEST START TEST Screen Selection of Single starts a test sequence for a single test with the duration of about 8 seconds Selection of CONT star...

Страница 11: ...he data Selection of FF allows the user to print a Report form or advance the paper on the internal printer Selection of STOP returns to the Test screen AUTO TEST CLOCK CONTRAST AUTO TEST Screen The A...

Страница 12: ...of current level then start at the 10A range and increase the current ranges until a valid reading appears After the current range is selected press the multi turn Knob again The following measuring...

Страница 13: ...val N Number of Tests desired A maximum of 10 minutes at 5 second intervals can be preset Measurement Notes Note 1 The instrument will halt the measuring process if excessive interference is detected...

Страница 14: ...Printer The internal thermal printer report has the following options Header Prints the header lines onto the report Data Prints the Data onto the report FF Form Feed The user may at the completion of...

Страница 15: ...vantage of this high accuracy the temperature of the device under test should be recorded The temperature probe is mounted to a flexible extension cable that plugs into the front panel of the MC2 This...

Страница 16: ...ment instruction step by step for the MC2 These are basic instructions simplified for the trained and qualified operator 1 Connect the test leads 2 Current 2 Potential to an un energized device to be...

Страница 17: ...ED 1 A warning message Rx Too High or will be displayed if the resistance being measured is too high for the range selected The is displayed when no input is measured on the P P terminals 2 The curren...

Страница 18: ...terials shall be the sole responsibility of purchaser This warranty shall not apply to any accessories parts or materials not manufactured or supplied by RAYTECH USA Inc Equipment must be returned pre...

Отзывы: