OPERATION:
8
–
0
AUTO (Automatic)
The AUTO (Automatic) mode will measure and store, automatically, numerous single sequential
tests at timed intervals that are preset by the operator. To preset the number of tests at timed
intervals please go to the Set-UP screen (See below) and select AUTO.
In this AUTO mode the instrument will continuously measure multiple tests until the operator
selects and presses STOP or until the number of tests that were preselected have been achieved.
The instrument will automatically save test results in a numbered memory location that will
increment by a timed interval.
AUTO Setup screen:
Ti = Test Interval
N = Number of Tests desired.
A maximum of 10 minutes at 5 second intervals can be preset.
Measurement Notes:
Note 1: The instrument will halt the measuring process if excessive interference is detected.
Note 2: If a warning message,”Rx Too High” is displayed, Or if ”+++.+” is displayed, Then
return to the “SET UP” screen and select the next higher current range.
The ”+++.+” is displayed when no input is measured on the +P &
- P terminals.
ARCHIVE:
9
–
0
The Archive or Memory allows over 2000 complete test results to be saved. The method used for
the memory location is:
Each memory location is designated a number, a date and time on which it was recorded, the
results measured, and the current it was tested at.
If multiple tests are taken in one test sequence the memory location will appear like this:
The test results can be printed from the Archive via the RS232 port or Parallel printer port.