20
RaySafe i3 Installation & service manual – System function tests
COMMUNICATION RANGE TEST
Follow the instructions below to verify the function of an environment with several Real-time Hubs installed.
NOTE:
If several systems are installed in the same building, make sure to keep distance between the hubs,
for good performance.
Check that the dosimeter communication range configuration is correctly tuned, that is, dosimeters currently not in
use does not appear on the display, but will appear when they approach the hub.
1. Position a dosimeter near Real-time Hub
A
.
2. Visually check that the dosimeter is shown on the screen connected to Real-time Hub
A
.
3. Move the dosimeter within range of Real-time Hub
B
.
4. Visually check that the dosimeter appears on the screen connected to Real-time Hub
B
within 30 seconds within
range.
5. Put the dosimeter in a rack and verify that it disappears from all displays within 9 minutes.
If anything fails in this test, see “Troubleshooting” on page 32.