13
RaySafe i3 Installation & service manual – Installation
CONNECTION OVERVIEW
Tech room
Hospital network
PoE
3 m Ethernet
30 m Ethernet
3 m Ethernet
HDMI cable
Hub
Video Unit
DVI video input
Power outlet
Lab
A2
A1
Ceiling
Figure 5. Cable connections. Purple parts provided by RaySafe. A1 and A2 are unconnected cable ends to
connect your computer to during the installation, for access to the configuration interface of the Video Unit and the
Real-time Hub.
NOTE:
Do not connect the Ethernet cables to the hospital network (cable end marked
A1
and
A2
5 on page 13) before you have configured the Video Unit (connection
A1
) and Real-time Hub
(connection
A2
).
MOUNT THE VIDEO UNIT
Place the Video Unit where suitable for your installation.
NOTE:
The Video Unit is a fanless computer. To avoid overheating, ensure good ventilation.
A
B C
Figure 6. The Video Unit uses three connections:
A
: power supply,
B
: Ethernet,
C
: HDMI (video output).
The Video Unit has a power button. When the unit is on, the LED is green. When the unit is manually turned off, by a
press on the power button, the LED is orange. Press the button to turn the unit on.