Set Diode Current
(
%
)
The set value must be less than 100% and above
the minimum current setting value, which
can be set to 0. If the set value is greater than
100, the default is input 100.
Send: ‘SDC 100\r’
Return: ‘SDC
:
100\r’
Other return values:
‘ERR: Input Err\r’
‘Laser is worked in AD Mode\r’
RCS
Read Current Setpoint
Send: ‘RCS\r’
Return: ‘RCS: 56.7\r’
(The current set value is 56.7 %)
RPRR
Read Pulse Repetition Rate (Hz)
Send: ‘RPRR\r’
Return: ‘RPRR:10\r’
(Repeat frequency is10Hz)
RBT
Read Board Temperature
Send: ‘RBT\r’
Return: ‘RBT:36.6\r’
RPW
Read Pulse Width (MS)
Send: ‘RPW\r’
Return: ‘RPW:5.5\r’
(pulse width is 5.5ms)
RCT
Read Laser Temperature
Send: ‘RCT\r’
Return: ‘RCT:34.5\r’
PERR
Reset Errors
Send: ‘PERR\r’
Return: ‘PERR\r’
SUT
Set Up Time (MS)
Send: ‘SUT 50\r’
Return: ‘SUT:50\r’
SDT
Set Down Time (MS)
Send: ‘SDT 50\r’
Return: ‘SDT:50\r’
RUT
Read Up Time (MS)
Send: ‘RUT \r’
Return: ‘RUT:50\r’
RDT
Read Down Time (MS)
Send: ‘RDT \r’
Return: ‘RDT:50\r’
PSRT
Program Start
Send: ‘PSRT 1\r’
Return: ‘PSRT:1\r’
PSTP
Program Stop
Send: ‘PSTP\r’
Return: ‘PSTP\r’