Manual Operation
R&S
®
UPV
166
Getting Started 1146.2078.62 ─ 09
●
Limits
A lower and/or upper limit can be defined for each measurement result.
If the limit is activated, each measured value is compared with the limit.
The set limits are shown in the bottom part of the combined display and appear as
a red limit marking on the bar graph.
As soon as one of the limits is exceeded, the measurement display and measure-
ment bar changes from green to red. If the measured value is within the limits, the
measurement display and measurement bar are displayed in green.
In order to monitor limit violations over a long period of time without having to
watch the measurement displays continuously, the lower and/or upper limit is dis-
played permanently in red as soon as it is exceeded. Pressing the START key
again or entering a new limit resets the limit to green.
The measurement display in the numeric display field always behaves in the same
way as the measurement display in the combined field; current limit violations can
therefore also be monitored in the numeric display field.
●
Extreme values
If this checkbox is activated, the lowest and highest measured value is displayed.
Display of the Min/Max values can be set separately for each measurement display
and for the two measurement channels.
The minimum and maximum values which have been measured are shown in the
top part of the combined display; an additional blue bar which records the span
between the extreme values is also displayed.
The Min/Max values can thus also be recorded over a long period of time without
the measurement displays having to be monitored continuously.
Measurement Displays