Getting Started
R&S
®
UPV
101
Getting Started 1146.2078.62 ─ 09
c) Next, the "Generator
Function" panel must be activated again; press the
WINDOW </> keys until the title bar of the panel is marked in blue.
Sine-wave signals are required; this setting is already selected.
To be able to measure the frequency response, the sine-wave signal must be
swept in the frequency. A fixed frequency of "1
kHz" is preset in the default
instrument setup.
To activate the sweep, select the "Sweep
Ctrl" field, open the parameter list
and select the desired setting.
d)
The parameter list closes and a number of additional fields appear in which fur-
ther sweep parameters can be set.
e) The "Next Step" field remains set to "Anlr Sync". This setting causes the next
sweep step to be triggered as soon as a valid measurement result is available
in the analyzer part. This synchronization optimizes the sweep speed.