Getting Started
R&S
®
UPV
97
Getting Started 1146.2078.62 ─ 09
3.3.9.1
Numeric Display Field
The numeric display field provides a clear overview of the numeric values from various
analyzer functions and displays a maximum of 8 numeric value results from the "Ana-
lyzer Function" panel. The individual columns are permanently assigned, from left to
right:
●
Column 1:
Measurement results of the function which is set in the
"Function" field of the "Ana-
lyzer
Function" panel.
●
Column 2:
Measurement results for the set "Level
Monitor" function.
●
Column 3: Measurement results for the "Input
Peak" function.
●
Column 4:
Display of either the frequency measurement result or, if "Freq
&
Phase" or
"Freq
&
GrpDel" is set, display of the frequency measurement result in the first row
and display of the phase or group delay measurement in the second row.
Combo Display
The combo display is used if the numeric value of a measurement result is to be sup-
plemented by analog bar graphs, limit monitoring or the storage of maximum and mini-
mum values.
The free selection and free scalability mean that especially important or critical results
can be monitored and tracked even when viewing the screen from a relatively long dis-
tance.
The easiest way to open combo displays is to highlight the respective measurement
display in the numeric display field (blue selection frame) and to select the display with
the ENTER key or a rotary
‑
knob click.
Brief Introduction to Operation