Bandwidth Extension settings
R&S
®
SMW-K555
34
User Manual 1179.5776.02 ─ 01
Remote Status ← RF port table
Indicates the remote connection status.
The RF port belongs to a secondary instrument; the remote connec-
tion to the secondary instrument can be established.
The remote connection is active during the time in that the correction
files are transferred to the secondary instruments.
The term
secondary
instrument describes a signal generator con-
nected in a primary-secondary mode and receiving the multi-instru-
ment trigger from the primary instrument (see "System Configuration
> Multi Instrument > Common Trigger > Multi Instrument Trigger =
Secondary").
and yellow
background
At least one of the secondary instruments cannot be reached, that is
the remote connection to it cannot be established.
Check the "Hostname / IP Address" of the secondary instrument and
select "Refresh" to check the status.
Remote command:
n.a.
4.2.3
Align RF ports configuration settings
Access:
1. Edit a bandwidth extension file as described in
Chapter 4.2, "Setup file editor set-
2. Select "Align RF Ports Config".
The dialog provides settings to align RF ports characteristics of the bandwidth
extension setup. Alignment includes phase corrections within the combined RF
output signal with extended bandwidth.
Settings:
Setup file editor settings