![R&S SMW-K555 Скачать руководство пользователя страница 24](http://html1.mh-extra.com/html/rands/smw-k555/smw-k555_user-manual_3082948024.webp)
Extending the bandwidth
R&S
®
SMW-K555
24
User Manual 1179.5776.02 ─ 01
If you change parameters of the RF output signal at the R&S
SMW, the RF ports are
not aligned anymore. For example, if you change the frequency or the level of the RF
output signal. Also, RF ports are not aligned if you change the bandwidth extension
setup file.
Align the RF ports again.
To align RF ports with a signal analyzer
The signal analyzer is switched-on and connected to the R&S
SMW via LAN, see
For prerequisites of the signal analyzer, see
1. Edit a bandwidth extension setup file.
See steps (1) and (3) in
"To manage bandwidth extension setup files"
2. Select "Align Config".
3. Select "Measurement Device" > "Analyzer".
4. Optionally, scan your network to add your signal analyzer.
5. If detected, select the signal analyzer in the "Instrument" selection.
The hostname of the signal analyzer is added automatically.
6. To add the signal analyzer manually, proceed as follows:
a) Select "Instrument" > "New".
b) Specify hostname or IP address of the signal analyzer.
7. Optionally, activate "Use Coupler", if you use a directional coupler to automate
alignment.
8. Activate the bandwidth extension setup configuration ("State > On").
The taskbar of the R&S
SMW displays the softkey "Bw Ext". The softkey indicates
that the RF ports are not aligned.
9. Select "Bw Ext".
10. Select "Align RF Ports".
Configuring the R&S
SMW