Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
357
User Manual 1173.9240.02 ─ 23
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<Offset>
Numeric value.
The unit depends on the scale of the x-axis.
*RST:
0
Manual operation:
See
CALCulate<n>:LIMit<k>:CONTrol:SHIFt
<Distance>
This command moves a complete limit line horizontally.
Compared to defining an offset, this command actually changes the limit line definition
points by the value you define.
Suffix:
<n>
.
irrelevant
<k>
Parameters:
<Distance>
Numeric value.
The unit depends on the scale of the x-axis.
Manual operation:
See
CALCulate<n>:LIMit<k>:CONTrol:SPACing
<InterpolMode>
This command selects linear or logarithmic interpolation for the calculation of limit lines
from one horizontal point to the next.
Suffix:
<n>
.
<k>
Parameters:
<InterpolMode>
LINear | LOGarithmic
*RST:
LIN
Example:
CALC:LIM:CONT:SPAC LIN
CALCulate<n>:LIMit<k>:LOWer[:DATA]
<LimitLinePoints>
This command defines the vertical definition points of a lower limit line.
Suffix:
<n>
.
irrelevant
<k>
Analyzing Results