![R&S FSW-K7 Скачать руководство пользователя страница 338](http://html1.mh-extra.com/html/rands/fsw-k7/fsw-k7_user-manual_3082493338.webp)
Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
338
User Manual 1173.9240.02 ─ 23
If the marker hasn't been active yet, the command first looks for the peak level in the
whole spectrogram and moves the marker vertically to the minimum level.
Suffix:
<n>
.
<m>
Usage:
Event
11.8.1.4
Marker Search Settings
The following commands define criteria for searches.
CALCulate<n>:MARKer<m>:PEXCursion
.........................................................................338
CALCulate<n>:MARKer<m>:PEXCursion
<Excursion>
This command defines the peak excursion (for
all
markers in
all
windows).
The peak excursion sets the requirements for a peak to be detected during a peak
search.
The unit depends on the measurement.
Application/Result display
Unit
Spectrum
dB
ADEMOD, RF
dB
ADEMOD, AM
PCT
ADEMOD, FM
kHz
ADEMOD, PM
RAD
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<Excursion>
The excursion is the distance to a trace maximum that must be
attained before a new maximum is recognized, or the distance to
a trace minimum that must be attained before a new minimum is
recognized
*RST:
5 PCT in AM displays, 50 kHz in FM displays, (0.5
RAD in PM displays)
Example:
CALC:MARK:PEXC 10dB
Defines peak excursion as 10 dB.
Manual operation:
See
Analyzing Results