Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
250
User Manual 1173.9240.02 ─ 23
Parameters:
<Mode>
ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST:
ABSolute
Example:
DISP:TRAC:Y:MODE REL
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:PDIVision
<Value>
This remote command determines the grid spacing on the Y-axis for all diagrams,
where possible.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Value>
numeric value WITHOUT UNIT (unit according to the result dis-
play)
Defines the range per division (total range = 10*<Value>)
*RST:
depends on the result display
Example:
DISP:TRAC:Y:PDIV 10
Sets the grid spacing to 10 units (e.g. dB) per division
Manual operation:
See
"Dev per Division/ Db per Division"
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
<Position>
This command defines the vertical position of the reference level on the display grid
(for all traces).
The R&S
FSW adjusts the scaling of the y-axis accordingly.
For measurements with the optional external generator control, the command defines
the position of the reference value.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Position>
*RST:
100 PCT = AF spectrum display; 50 PCT = time
display
Example:
DISP:TRAC:Y:RPOS 50PCT
Configuring the Measurement