Analysis
R&S
®
FSW-K7
153
User Manual 1173.9240.02 ─ 23
If a positive offset is entered, the markers T1 and T2 are placed below the active refer-
ence point. If a negative value is entered (for example for notch filter measurements),
the markers T1 and T2 are placed above the active reference point.
Figure 6-2: n dB down marker function
The following marker function results are displayed:
Table 6-1: n dB down marker function results
Label
Description
M1
Current position and level of marker 1
ndB
Offset value (n dB down)
ndB down Bw /
PWid
Determined bandwidth or pulse width (zero span) at the offset
Q-factor
Quality factor of the determined bandwidth (characteristic of damping or resonance)
T1, T2
Current position and level of the temporary markers
If the required position for the temporary markers cannot be determined uniquely, for
example due to noise, dashes are displayed as a result.
Working with Markers in the R&S FSW Analog Demodulation application